18
6. Preparing for pumping
For chapter 6.2 Mains operation:
l
Use only the power cord that comes
with the Symphony
.
l
Make sure the voltage of the power
cord is compatible with the power
source.
Warning
l
Use the pump only with the appropriate
program card (Symphony Standard or
Symphony PLUS).
Optional for pump with battery:
l
The built-in battery must be charged for
12 hours on receipt of the device.
l
The pump can be used during charging.
l
If the pump is to remain unused for a
long time, charge the battery every two
months (for 12 hours each time).
l
The optimum service life of the battery
is ensured if it is not fully discharged
on a regular basis or if the pump is
charged while not in use.
Info
i
1
6.1 Card and cord protector
To remove, unscrew the card
and cord protector and slide it
out of the guide rail.
1
Insert the power cord into the
power socket on the back of
the pump.
6.2 Mains operation
Pumping time
(fully-charged battery)
approx. 60 min.
Charging time
(for full charge)
12 hours