9. Troubleshooting
Problem
Solution
If the motor
is not running
Check whether mains power/battery is available.
Check that the device is switched on.
Check whether the program card is inserted correctly in the card
compartment on the back of the breast pump.
If there is low
or no suction
Assemble the pump set in accordance with the instructions in chapter
6.4.
Check that all connections of the pump set and breast pump are
secure.
Express in accordance with chapter 7. Make sure that the breast
shield forms a complete seal around the breast.
See chapter 6.4 „Assembling the pump set“.
l
The ends of the tubing must be firmly connected to the connector
lid and the opening in the Symphony cap.
l
The Symphony cap must be firmly fitted to the breast pump
membrane to form a seal.
l
The membrane must be intact.
l
The membrane must be correctly placed into the connector and
the connector lid properly closed.
l
Do not kink or trap the tubing during expression.
l
All components must be completely dry.
9.1 Troubleshooting