Service manual P9-0200-S
25 of 82
Version 2.0.2/2012
Medema Production A/S
Troubleshooting
Trouble
Possible causes
Remedy
MC Concept will not
run.
There is no light in
the battery indicator.
The MC Concept is not
turned on (I/O button)
The batteries are
discharged.
Thermal protection
popped out.
Charging connector sits
in.
Malfunction.
Turn the MC Concept on I/O
button).
Oplad batterierne i mindst 12
timer.
Press thermal protection back
in again.
Turn off the charger and remove
the charging socket.
Contact supplier.
Mini Crosse MC
Concept will not
run, but there is
light in the battery
indicator.
A function key for a
function that is not
mounted is pressed.
Indicator light will be lit
and there will be a light for
the activated function key.
Fault in the electronics.
Press the funtionstast that lights
and turn off at I / O button.
Wait 5 seconds., Press I / O
button again.
Contact supplier.
The driving speed is
too low
Not enough air in the tires.
The wheelchair is
overloaded.
The seat is raised above
its lowest position. (Light
above I / O button is
shown).
Battery capacity is too low.
Pump pneumatic wheels to 2.8
bar.
Stop and wait a few minutes
before starting again.
Lower the seat.
Recharge or replace batteries.
Too few km per
charging.
Batteries are not working.
The charger is not
working.
Not enough air in the tires.
The method of charging is
wrong.
Defective gear motor
(indicator lamp lights on
the control panel below
the I / O button)
Recharge your batteries and
make sure that the green light
on the charger lights.
Contact supplier.
Pump pneumatic wheels to 2.8
bar.
Read the manual.
Contact supplier.