Turn ON mode
This function enables the device to turn on automatically at a predetermined time. The device will automatically turn on and commence a sleep
oximetry test (this test is also capable of monitoring the patient all day long, and includes as a step counter and a triaxial accelerometer to measure
the VMU)
The device will turn off automatically at a predetermined time.
WARNING
If the automatic turn-ON function is set it will be impossible to turn off the device during
the test. The closed lock icon located on the top center of the screen warns the user of
the current setting.
Select the item by touching the display and choose between the following options:
Manual
Automatic
Manual turn-ON: allows to set the turning on of the device by using the specific ON/OFF key.
Automatic turn-on: enables the programming of the frequency and duration the device turns on. Select the desired item and press OK. If the user
selects automatic turn-on she/he may choose from the following options:
One time only
Once a week
Monday to Friday
Saturday-Sunday
Every day
For each option a menu appears to set the day and time that the device will turn on and off.
If turn ON mode was previously enabled, and the user turns on the device at a different time
from that of the programmed time, the device will show the screen on the right. To continue
touch the OK icon and insert the password 1 2 2 3 3 3; by touching the
the device will
turn off.
Oximetry setup
When entering the Oximetry Setting menu the following items are shown:
Alarms Setting
SpO2 Sampling Rate
Pulse tone ON/OFF
Default alarms
Alarms setting
Access to this function allows to setup reference values for SpO2 and BPM; an acoustic alarm will warn the user if the SpO2 and BPM during a test
fall below the minimum threshold or rise above the maximum threshold of the SpO2 and BPM values previously set.
The first configurable parameter is the sound: it is possible to select the type of sound and
volume; both parameters have two values:
Type of sound
1 kHz
4 kHz
Volume
min
MAX
As can be seen in the image on the right.
Touch the corresponding squares/panes to select the desired values.
By touching OK in the bottom right side the user will access the minimum and maximum
threshold of the two oximetry parameters. For each parameter the screen allows to set the
alarm on or off. (By touching the ON and OFF icons) or to change the threshold value with
the icon
.