spirodoc
SPIRODOC
cod. 980156
Rev 2.4
EN 20/46
User manual
The sequence of values is the following:
Parameter
Min value setting
Max value setting
SpO2 min
60
99
SpO2 max
60
99
BPM min
60
120
BPM max
60
120
WARNING
If the maximum value of a SpO2/BPM parameter is set lower or equal to the minimum value the setting will not take effect. The device
will emit an acoustic warning and automatically return to the setting of the minimum value.
After having set the maximum BPM value touch OK to set the acoustic alarms. The sequence
is the following:
Finger not inserted
Sensor not inserted
Battery discharged
The image on the right is an example of the on-off alarm for the finger being used with the oximetry sensor.
SpO2 sampling rate
This function allows to set the time that elapses between the sampling/recording of two consecutive oximetry readings; touch one of the two
visualized icons: 2 seconds or 4 seconds, then touch OK to set the selected value and the device will automatically return to the service menu.
Pulse tone ON/FF
This setting enables the heart rate tone (beep) during oximetry testing
WARNING
The heart rate tone (beep) is always disabled during sleep oximetry testing
(please view paragraph 3.13.2).
Default alarms
This function allows to restore all default settings.
To confirm touch the “YES” icon. The
settings return to the standard values. The image on the right shows the factory default
settings.
After a few seconds the device will return to the oximetry setting screen.
Select predicted
Select the item by touching the display.
A list of predicted values is shown; select the Predicted value desired.
Adult
Pediatric
ERS
Knudson
Knudson
Knudson
USA
Knudson
ERS
Zapletal
MC-Barcelona
Zapletal
JRS
Knudson
Select the pair to use and touch OK. The Predicted values are set and the device returns to the Service Menu.
Select standard
Select the item by toughing the display.
Select the standard to be used (ATS/ERS, or NHANES III) and touch OK, the setting takes effect and the device returns to the Service Menu.
WARNING
If the NHANES III standard is selected it is not possible to set or modify the predicted values.
Set parameters
It is possible to select the type of calculated parameters during spirometry and oximetry testing. For each of the two categories the user may select
from the following three options:
simplified
personal