Connect the sensor to the device: insert the connector with the arrow (printed on the
connector) face-up, as shown:
Choose a high perfusion site, easily adaptable to the sensor.
Insert the finger into the sensor until the finger touches the end of the probe. Ensure that
the bottom part of the finger completely covers the detector. If the finger cannot be placed
properly inside the sensor try another finger.
Place the sensor so that the cable rests on the back of the hand. This ensures that the
light source rests. On the side of the nail and the reader on the lower part of the hand.
Select one of the tests that can be performed with
SPIRODOC
To access the oximetry area touch the
icon on the main screen and subsequently select the type of oximetry test to carry out.
SpO2/BPM spot test
Sleep oximetry test (ODI)
Six minute walk test
If the following message appears upon start-up:
WARNING OXIMETER NOT PRESENT
This means that your device does not have this function.
If instead the following message appears:
WARNING THE OXIMETER IS NOT ENABLED
This means that the oximetry function is included, however the internal application has yet to be enabled. In this case please contact a service
center or the manufacturer.
WARNING
Before carrying out a test, if the power supply value is low the following message will appear:
Low battery level
Touch the ESC icon to exit the test, otherwise after a seconds will start the test.
In the event that a test is interrupted due to a complete battery discharge, the next time the device is turned on the following message is
displayed:
WARNING
Wrong interruption of last oximetry test
At the same time an intermittent beep is emitted for 4 seconds.
Subsequently the SPIRODOC returns to the main screen.
WARNING
Avoid twisting the sensor’s cable as this may compromise measurement accuracy and the integrity of the sensor itself, also do not apply
excessive force when using, connecting, disconnecting or storing the oximetry sensor.
The first few seconds are used to find the best signal possible; after which the
SPIRODOC
timer resets itself and the device starts recording data.
For any type of oximetry test if the sensor is not properly connected the following message will be displayed on screen after a few seconds:
WARNING
Sensor unplugged
At the same time
SPIRODOC
emits an acoustic alarm (if previously set in the service menu).
If the sensor has been connected properly but the finger has not been properly inserted in the sensor the following message will be displayed on
screen.
WARNING
Finger not inserted
At the same time
SPIRODOC
emits an acoustic alarm (if previously set in the service menu).
If the signal reaches the sensor properly, after a few seconds the device will emit an acoustic signal while also displaying the values on screen.
The alarms can be customized, the procedure is described in paragraph 3.6.