background image

Technical specifications for CCC

220

 

 

 

CCC

220 

 

Charging current (ripple-free): 

20 A 

Compatible batteries: 

 

12 V, 40 Ah .. 300 Ah – Type Gel, AGM 

Dimensions, B x H x D (incl. handle): 

220 mm x 75 mm x 205 mm 

Weight: 

 

 

2.6 kg 

 

Efficiency: 

 

 

not less than 90% 

Short-circuit and faulty polarization safe 

 

Mains power:   

 

230V 50Hz 

 

 - labeled in accordance with:  

the low voltage Directive and the EMC directive

 

Enclosure rating: 

Splash proof, dustproof and water proof 

 

IP54,IP65 and IP67 

Insulation class:  

 

II (double-insulated) 

 

 

 

 

This means that the charger can be connected to an ordinary 

 

 

 

power socket without earth connection. 

 

Ambient environment 

Temperature 

Relative air humidity 

Air pressure 

Use 1) 

-10°C .. +30°C  

10% .. 90% 

70 kPa .. 106 kPa 

Transport and storage 2)  -40°C .. +70°C 

10% .. 90% 

70 kPa .. 106 kPa 

1): Bearing in mind the safety information, page 2 
2): In accordance with the packaging's label 

 

 

 

Warning:

  

The charger may only be used for charging rechargeable 12

V lead batteries

  

Warning:

  

You must not recharge batteries that are not rechargeable! 

 

Energy transfer from the mains to the battery uses a patented power circuit. 
The charging process is controlled by a built-in microcomputer. The current state of the battery is adapted 
automatically. This will optimize the lifetime of the battery. 
Once the battery is fully charged, the status changes to trickle charging. Overcharging is not possible. 
The charging time is the shortest possible. 

 

The "CHARGING", "COMPLETED" and "ERROR" lights on the front of the charger signal the charging 
status. 
Any faults in the battery or charger are signaled using error codes (see diagram 2). 

 

Option for (program options): 

 
• TCS additional charging (patented):                                    

 

• Temperature-compensated standby charging 

If the charger has temperature-compensated standby charging, it must be placed in the same 
room (the same temperature) as the wheelchair if batteries need to be charged! 

• Controlled charging of significantly depleted batteries 
• Display of charged capacity (Ampere-hour meter) 
• Adaption of charging characteristics to current requirements. 

 

Important:  

The charger's properties (program options) may only be changed by authorized personnel! 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

6

 

Summary of Contents for CCC220

Page 1: ...battery charger CCC220 12V 20A Splash proof dustproof and water proof IP54 IP65 and IP67 Manufacturer Tel 45 4717 1900 E mail JK Medico JK Medico DK J K Medico ApS Dam Enge 30 DK 3660 Stenloese Denmark WWW JK Medico DK ...

Page 2: ... on 4 Turn the charger off by removing the charger s power plug from the outlet 6 If it becomes necessary to interrupt charging before the COMPLETED light comes on turn the charger off by removing its power plug from the outlet before removing the charging plug from the wheelchair s charging socket This will prevent sparks and unnecessary wear to the charging plug 1 The charger tests itself when o...

Page 3: ...Warning For safety reasons batteries must not be repeatedly recharged if the charger has reported that the battery is faulty see diagram 2 Contact your battery supplier Warning For safety reasons repeated charging should not be undertaken using a charger that has reported Fault in charger see diagram 2 Contact your charger supplier for repairs Warning The charger s base and rear plate cooling plat...

Page 4: ...s that signal the charger s properties set before delivery 1st flash TCS additional charging patented 2nd flash Temperature compensated standby charging 3rd flash Charging of batteries with off load voltage below 16 8V possible 4th flash Ampere hour meter Red flash ERROR light Property passive Green flash COMPLETED light Property active Figure 1 Signaling of the charger s configuration when the ch...

Page 5: ...apacity Contact battery supplier 3 4 OFF ON 5 flashes Battery fault Battery voltage too high System fault 2 External error source Group flash with ERROR OFF OFF 6 flashes Charger temperature too high Check that the charger is positioned so that there is 5 cm of free space on all sides OFF OFF ON No battery connected Standby mode 5 Figure 2 Signaling of faults 1 Always possible charging will recomm...

Page 6: ... Figure 2 Display showing charged capacity when the charger are disconnected from the battery once charging has ended 5 Number of yellow flashes Number of green flashes Charged capacity ampere hours 3 yellow flashes 2 green flashes 32 ampere hours Ah ...

Page 7: ...charge batteries that are not rechargeable Energy transfer from the mains to the battery uses a patented power circuit The charging process is controlled by a built in microcomputer The current state of the battery is adapted automatically This will optimize the lifetime of the battery Once the battery is fully charged the status changes to trickle charging Overcharging is not possible The chargin...

Page 8: ... least 5 cm free space on all sides Position the charger so that the control lights are visible Do not place the charger in direct sunlight by radiators or any other heat source Important Avoid covering the charger in any way Dismantling and repairs Warning For safety reasons the charger must be dismantled and or repaired by authorized personnel only Contact your charger supplier if required for i...