Test Principle
The blood sample is drawn into the test strip's reaction chamber through capillary action. Glucose in the
sample reacts with glucose dehydrogenase and mediator in the test strip. This reaction creates electrical
currents. The electrical currents produced are proportional to the glucose concentration in the blood
and converted to the equivalent glucose concentration values calculated based on the algorithm
programmed in GlucoDr auto Meter.
- The GlucoDr auto Meter is designed for use with GlucoDr auto Test Strips
(manufactured by
All Medicus Co., Ltd.
) only.
- The GlucoDr auto Blood Glucose Monitoring System is calibrated to produce values equivalent to results
on plasma specimens obtained from a laboratory analyzer (YSI 2300 STAT Plus).
- The GlucoDr auto Blood Glucose Monitoring System should not be used for the diagnosis of diabetes.
Precautions for Use
For Self-testing:
GlucoDr auto Blood Glucose Monitoring System is for quantitative blood glucose testing
in fresh capillary whole blood drawn from the
fi
ngertips, palm, arm, thigh, and calf. The GlucoDr auto Blood
Glucose Monitoring System is for testing outside the body (
in vitro
diagnostic use). The GlucoDr auto Blood
Glucose Monitoring System is intended for use by diabetic lay users for self testing to improve management
of their blood glucose level. The GlucoDr auto Blood Glucose Monitoring System should not be used for the
diagnosis of or screening of diabetes mellitus, or for neonatal use.
For professional use:
GlucoDr auto Blood Glucose Monitoring System is for quantitative blood glucose
testing in venous, arterial whole blood and fresh capillary whole blood drawn from the
fi
ngertips, palm,
arm, thigh, and calf. The GlucoDr auto Blood Glucose Monitoring System is for testing outside the body (
in
vitro
diagnostic use). The GlucoDr auto Blood Glucose Monitoring System is intended for use by healthcare
professionals to improve management of patient’s blood glucose level. The GlucoDr auto Blood Glucose
Monitoring System should not be used for the diagnosis of or screening of diabetes mellitus, or for
neonatal use.
Intended Use
*
PRECAUTIONS