TREADMILLS MOD. 870A/870S/870C
Rev.1/01.07.2019
© Medisoft RAM Italia S.r.l.. All rights reserved.
24
3.6
VERIFICATION OF SPEED PARAMETERS
WARNING!!
IMPORTANT!!
Perform this procedure before using the system, otherwise you could
create dangerous situations for the user.
Apply a piece of scotch on running belt and another on the treadmill side cover
(the latter used as a reference point). Count visually the number of times that
the scotch applied to the running belt will overcome the scotch of the side cover
over an accurate minute time.
Check if the counted values is the same of the tables below.
Treadmill mod. 870 A - Clinical
Units for the speed
chosen
Speed value to be set
Expected count
Mph
3
27 ± 1
Km/h
3
17 ± 1
Tappeto rotante mod. 870 S - Sport
Units for the speed
chosen
Speed value to be set
Expected count
Mph
3
23 ± 1
Km/h
3
15 ± 1
Tappeto rotante mod. 870 C - Athlete
Units for the speed
chosen
Speed value to be set
Expected count
Mph
3
21 ± 1
Km/h
3
14 ± 1
If the values counted are different from the count expected please contact
the Service department of Medisoft RAM Italia to verify the settings.
Note
: This module is also provided on paper attached to the installation
procedure.