TREADMILLS MOD. 870A/870S/870C
Rev.1/01.07.2019
© Medisoft RAM Italia S.r.l.. All rights reserved.
39
ANNEX D
– Troubleshooting
Problem
Possible Cause
Remedy
Cannot switch
treadmill on
.
Power cable not connected
Connect the power cable to
the power source
The RED Stop Switch is
activated
Deactivate the Red Stop
Switch (Turn the switch
clockwise, as indicated on
the switch surface)
Mains power failure
Check mains power with
another appliance
Power cable damaged
Repair or replace the power
cable
Fuses blown out
Replace the fuses as
described in section 5.4
Accidental discontinuity of
the connection between
the Control Panel and the
Motor space
Check the connection as
described in section 3.2
(see Fig. 6)
Electronic fault in the
Control Panel
Contact Medisoft RAM
Italia service
Treadmill makes
unexpected noise
The treadmill is not placed
on a adequate surface
Move the treadmill to
appropriate position
Static converter or motor
damaged
Contact Medisoft RAM
Italia service
Gear motor damaged
Contact Medisoft RAM
Italia service
Roller bearings damaged
Contact Medisoft RAM
Italia service
Drive belt bearings
damaged
Contact Medisoft RAM
Italia service
The treadmill
stops working,
Blockage of treadmill due
to excess load.
Let the system rest and
then try again.