8
2.
Press M to mark this test as a control solution test
With “
QC
” displayed, the meter will store your test result in
memory under
“
QC
”. If you press
M
again, the “
QC
” will
disappear and this test is no longer a control solution test.
Warning:
When doing the control solution test, you have to mark it so that the test
results will
NOT
mix with the blood glucose
TEST RESULTS
stored in the
memory. Failure to do so will mix up the blood glucose test results with the
control solution test results in memory.
3.
Apply control solution (g)
Shake the control solution vial thoroughly before use. Squeeze
out the first drop and wipe it off, then squeeze out another drop
and place it on the tip of the vial cap. Hold the meter to move the
absorbent hole of the test strip to touch the drop.
Once the
confirmation window fills completely, the meter will begin
counting down.
NOTE:
To avoid contaminating the control solution, do not directly apply control
solution onto a strip.
4.
Read and compare the result
After counting down to 0, the control solution test result will
appear on the display. Compare this result with the range printed
on the test strip vial and it should fall within this range. If not,
please read the instructions again and repeat the control
solution test.