24
For further information, please refer to the test strip package insert.
Important Control Solution Information
⚫
Use only our control solutions with your meter.
⚫
Do not use the control solution beyond the expiration date or 3
months after first opening. Write the opening date on the control
solution vial and discard the remaining solution after 3 months.
⚫
It is recommended that the control solution test be done at room
temperature 20°C to 25°C (68°F to 77°F). Make sure your control
solution, meter, and test strips are at this specified temperature
range before testing.
⚫
Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
⚫
Store the control solution tightly closed at temperatures between
2°C to 30°C (35.6°F to 86°F). Do
NOT
freeze.