6.4 Belt vibrations
Indication
Treadmill running belt vibrates
*
in walking speeds which varies from
0.1-1.2 kmh.
* Fast and small forward & backward movements with a duration
approximately of 10 times per second.
Reason for error/fault 1.
Inner side of the treadmill belt is very dry, which causing a high friction
between the running board and the treadmill inner surface.
2.
Extreme high treadmill belt tension, which caused by unappropriated
tension of the back pulley.
Corrective actions
Solution 1
Lubricate the running board surface bellow the running belt with
Silicon Lubricant Liquid, run the treadmill on high speed so the
liquid will spread uniformly all over the overlapped surface, then test
the belt vibration again for the low speeds.
Solution 2
If Solution 1 not solves the problem, there is a need to modify front
motor gains in the servo driver. For this, please announce
MediTouch and schedule a remote session with MediTouch tech
team.