Page 26 of 34
www.meditouch.co.il
info@meditouch.co.il
Tel: (972)-9-8637477
Fax: (972)-9-8852935
45 Hamelacha, Poleg industrial zone, Netanya, Israel.
Zip: 4250574, PO Box: 8306
DO-15-01-12
BalanceTutor service manual 211102.doc
6.8 No display at patient screen
Indication
Messages
on patient screen “Check Signal Cable”
Reason for error/fault Screen not configured, Damaged cable/plugs during assembly, HDMI plug
not fully inserted to PC HDMI socket
Corrective actions
Visual inspection on impurity of DVI-HDMI cable, Screen setup, check
screen and pc connectors.
6.8.1 Cable plugs not fully inserted/connected
Indication
HDMI plug not fully inserted to PC HDMI socket
Corrective actions
Configure the screen using screen control buttons to direct the video to the
DVI port.
A.
Remove electrical board cover
B.
Confirm HDMI Plug it properly connected to the PC HDMI socket
C.
Restart the system and press channel selection on the patient screen
to detect the DVI connection channel
Note: on the same principle, check the DVI plug connected to the patient
screen
Risk from electrical shock. Risk of damaging electrical internal components.