853591-1_AIR-I-MTINI-1 Avant AIR+ Installation Manual

 

13 

Effective Date: 12-08-2017 

 

EMC Precautions

 

The Avant Audiometer needs special precautions regarding EMC and needs to be installed and 

put into service according to the following EMC information. 
 

List of all cables and maximum lengths of cables, transducers and accessories: 

 

Transducer / Accessories 

Maximum Cable length 

USB Cable 

3 meters 

Insert Earphones 

2 meters 

All Headsets 

2 meters 

All Microphones 

2 meters 

  

Warnings! 

 

 

The use of accessories, transducers and cables other than those specified, with the 
exception of transducers and cables sold by the manufacturer of the Avant Audiometer 

as replacement parts for internal components, may result in increased emissions or 

decreased immunity of the Avant Audiometer. 

 

The Avant Audiometer should not be used adjacent to or stacked with other equipment 
and if adjacent or stacked use is necessary, the Avant Audiometer should be observed 

to verify normal operation in the configuration in which it will be used. 

 

The Avant Audiometer may be interfered with by other equipment, even if that other 
equipment complies with CISPR emission requirements. 

 

The Avant Audiometer does not have life supporting function 

 

Portable and mobile RF communications equipment can affect the Avant Audiometer. 

 

 

Guidance and manufacturer’s declaration – electromagnetic emissions 

The Avant Audiometer is intended for use in electromagnetic environment specific below. The customer or the user of 

the Avant Audiometer should assure that it is used in such an environment. 

Emission test 

Compliance 

Electromagnetic environment - guidance 

RF emissions 

CISPR 11 

Group 1 

The Avant Audiometer uses RF energy only for 

its internal function. Therefore, its RF emissions 

are very low and 

are not likely to cause any interference in nearby 

electronic equipment. 

 

RF emissions 

CISPR 11 

Class B 

The Avant Audiometer is suitable for use in all 

establishments, including domestic 

establishments and those directly connected to 

the public low - voltage power supply network 

that supplies buildings used for domestic 

purposes. 

Harmonic emissions 

IEC 61000-3-2 

Non applicable 

Voltage fluctuations / flicker emissions 

IEC 61000-3-3 

Non applicable 

 

 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The Avant Audiometer is intended for use in electromagnetic environment specific below. The customer or the user of 

the Avant Audiometer should assure that it is used in such an environment.

 

Immunity test 

IEC 60601 

test level 

Compliance level 

Electromagnetic 

environment - guidance 

Electrostatic discharge 

(ESD) 

 

IEC 61000-4-2 

+/- 6 kV contact 

+/- 8 kV air 

+/- 6 kV contact 

+/- 8 kV air 

Floors should be wood, 

concrete or ceramic tile. If 

floors are covered with 

synthetic material, the 

relative humidity should be 

at least 30%. 

Summary of Contents for AVANT AIR PLUS

Page 1: ...www medrx int com MANUAL INSTALLATION AVANT AIR AScreening Audiometer...

Page 2: ...Cleaning 17 Limited Warranty 18 www medrx int com Manufacturer MedRx Inc 1200 Starkey Rd 105 Largo FL 33771 U S A Toll Free 888 392 1234 727 584 9600 Fax 727 584 9602 E mail medrx medrx usa com www me...

Page 3: ...or training in the field of audiometry Indication for Use Statement This device is an audiometer For use by professionals with education and or training in the field of audiometry to conduct diagnost...

Page 4: ...M 5 GB free hard drive space Available 2 0 USB port Windows 7 or 8 Professional 32 or 64 bit MedRx Recommended Computer Specs Windows PC compatible computer Intel Dual Core 2 0 GHz or better 3 GB RAM...

Page 5: ...MTINI 1 Avant AIR Installation Manual 5 Effective Date 12 08 2017 Avant AIR Bottom View with Connectors Transducers and Accessories Talkback Mic Patient Response Switch Operator Mic and Monitor may va...

Page 6: ...vary depending on the USB port selected Consult your computer s documentation 2 Double Click on Setup to launch If requested Accept Permission to Install Software 3 On the Setup screen choose Install...

Page 7: ...Agreement select I accept click Next 6 This screen sets the language and location choice Make a selection and click Next 7 This screen indicates the location of the program files The default location...

Page 8: ...nt AIR Installation Manual Effective Date 12 08 2017 8 Installation of Program is ready to start To continue click Install To make changes click Back 9 Installation is in process 10 When the installat...

Page 9: ...Windows drivers When this screen appears and all devices are Ready to use click Close Next you need to confirm or set the Default Windows sound card settings This will route all non AVANTTM Windows so...

Page 10: ...specific Avant Audiometer will appear in both Output and Input Device Tab Click Audio Properties In Windows Sound control panel make sure the MedRx Audio Device is not set as default If it is default...

Page 11: ...ng Calibration Files Each Avant AIR is calibrated in compliance with the ANSI S3 6 standard This calibration procedure results in a series of files that the Audiometer software reads to keep the hardw...

Page 12: ...vant AIR Installation Manual Effective Date 12 08 2017 2 After a few seconds this message will appear Click Load 3 When the files are finished being loaded this message will appear Click YES to comple...

Page 13: ...meter Guidance and manufacturer s declaration electromagnetic emissions The Avant Audiometer is intended for use in electromagnetic environment specific below The customer or the user of the Avant Aud...

Page 14: ...P 80 to 800 MHz d 2 33 P 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation dist...

Page 15: ...ximum output power of transmitter W Separation distance according to frequency of transmitter meters 150 kHz to 80 MHz d 1 17 P 80 MHz to 800 MHz d 1 17 P 800 MHz to 2 5 GHz d 2 33 P 0 01 0 12 0 12 0...

Page 16: ...ll components with patient contact are made of bio compatible materials This device does not produce any adverse physiological effects Install the device as directed by this manual to achieve optimal...

Page 17: ...tips are single use components and should not be re used by another patient 2 It is recommended that 70 Isopropyl Alcohol should be applied to a soft clean cloth or tissue not directly on the componen...

Page 18: ...he system is two years old or less and has not been misused abused or damaged Such damage includes but is not limited to dropping exposure to excessive heat greater than 37 78 C and water liquid damag...

Reviews: