background image

853591-1_AIR-I-MTINI-1 Avant AIR+ Installation Manual 

Effective Date: 12-08-2017 

Software Installation 

 

 

 

1.

 

Insert the AVANT 

Audiometer

 USB Flash 

Drive into USB port: 

 

Launch 

My Computer

  

 

Locate USB Drive 

 

(E:) may vary depending on the USB port selected.

 

Consult your computer’s documentation.

 

 

 

 

2.

 

Double Click on 

Setup 

to launch: 

 

If requested, Accept Permission 

to Install Software  

 

 

 

 

3.

 

On the Setup screen, choose 

Install 

AVANT Audiometer

 

 

NOTE

: No MedRx driver installation is 

required with the AVANT Audiometers.  

 

Also, the Electronic Copy of this Manual is 
located under 

Manual. 

 

 

 

 

 

 

4.

 

This is the Welcome screen. 

 

To continue, Click 

Next

.

 

 

Summary of Contents for AVANT AIR PLUS

Page 1: ...www medrx int com MANUAL INSTALLATION AVANT AIR AScreening Audiometer...

Page 2: ...Cleaning 17 Limited Warranty 18 www medrx int com Manufacturer MedRx Inc 1200 Starkey Rd 105 Largo FL 33771 U S A Toll Free 888 392 1234 727 584 9600 Fax 727 584 9602 E mail medrx medrx usa com www me...

Page 3: ...or training in the field of audiometry Indication for Use Statement This device is an audiometer For use by professionals with education and or training in the field of audiometry to conduct diagnost...

Page 4: ...M 5 GB free hard drive space Available 2 0 USB port Windows 7 or 8 Professional 32 or 64 bit MedRx Recommended Computer Specs Windows PC compatible computer Intel Dual Core 2 0 GHz or better 3 GB RAM...

Page 5: ...MTINI 1 Avant AIR Installation Manual 5 Effective Date 12 08 2017 Avant AIR Bottom View with Connectors Transducers and Accessories Talkback Mic Patient Response Switch Operator Mic and Monitor may va...

Page 6: ...vary depending on the USB port selected Consult your computer s documentation 2 Double Click on Setup to launch If requested Accept Permission to Install Software 3 On the Setup screen choose Install...

Page 7: ...Agreement select I accept click Next 6 This screen sets the language and location choice Make a selection and click Next 7 This screen indicates the location of the program files The default location...

Page 8: ...nt AIR Installation Manual Effective Date 12 08 2017 8 Installation of Program is ready to start To continue click Install To make changes click Back 9 Installation is in process 10 When the installat...

Page 9: ...Windows drivers When this screen appears and all devices are Ready to use click Close Next you need to confirm or set the Default Windows sound card settings This will route all non AVANTTM Windows so...

Page 10: ...specific Avant Audiometer will appear in both Output and Input Device Tab Click Audio Properties In Windows Sound control panel make sure the MedRx Audio Device is not set as default If it is default...

Page 11: ...ng Calibration Files Each Avant AIR is calibrated in compliance with the ANSI S3 6 standard This calibration procedure results in a series of files that the Audiometer software reads to keep the hardw...

Page 12: ...vant AIR Installation Manual Effective Date 12 08 2017 2 After a few seconds this message will appear Click Load 3 When the files are finished being loaded this message will appear Click YES to comple...

Page 13: ...meter Guidance and manufacturer s declaration electromagnetic emissions The Avant Audiometer is intended for use in electromagnetic environment specific below The customer or the user of the Avant Aud...

Page 14: ...P 80 to 800 MHz d 2 33 P 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation dist...

Page 15: ...ximum output power of transmitter W Separation distance according to frequency of transmitter meters 150 kHz to 80 MHz d 1 17 P 80 MHz to 800 MHz d 1 17 P 800 MHz to 2 5 GHz d 2 33 P 0 01 0 12 0 12 0...

Page 16: ...ll components with patient contact are made of bio compatible materials This device does not produce any adverse physiological effects Install the device as directed by this manual to achieve optimal...

Page 17: ...tips are single use components and should not be re used by another patient 2 It is recommended that 70 Isopropyl Alcohol should be applied to a soft clean cloth or tissue not directly on the componen...

Page 18: ...he system is two years old or less and has not been misused abused or damaged Such damage includes but is not limited to dropping exposure to excessive heat greater than 37 78 C and water liquid damag...

Reviews: