14
ARC-I-MARCI-2 MedRx AVANT ARC Installation Manual
Effective Date: 10-21-2016
Loading Calibration Files
Each AVANT
ARC
audiometer is calibrated in
compliance with the ANSI S3.6 standard. This
calibration procedure results in a series of files
that the AVANT
ARC
software reads to keep the
hardware in calibration. These files are supplied
on a CD bearing the same serial number as your
AVANT
ARC
device.
The final step before using your AVANT
ARC
to
evaluate hearing is to load these device-specific
calibration files onto the computer used to
operate the AVANT
ARC
device.
1. With the Calibration CD in the drive, open
the AVANT
ARC
software and click:
•
Function.
•
Calibrate.
•
Load Calibration.
2. After a few seconds, the CD will “spin up”
and this message will appear.
•
Click
Load.
3. When the files are finished being loaded,
this message will appear:
•
Click
YES
to complete loading the
calibration.