ARC-I-MARCI-2 MedRx AVANT ARC Installation Manual
19
Effective Date: 10-21-2016
Symbols that may be used:
or SN
Read the instruction manuals
for safe usage of the device.
(operating instructions)
Indicates that the device serial
number will follow
Type B applied part. (Type B
equipment)
Manufacturer (MedRx)
Authorized Representative in
Europe
Non-ionizing electromagnetic
radiation
Special Disposal Required
Humidity
Limitation
Caution, General warning sign
Temperature limitation
Read the instruction manuals for
safe usage of the device
(operating instructions)
Class II equipment
Start (of action)
Stop (of action)
Percentile Setup
Calibration
Loudspeaker (Speaker)
Headphones
Microphone
Recording
Cleaning
- Recommended Procedures for Cleaning & Disinfection
1. Foam ear tips and probe tubes are single use components and should not be re-used on another patient.
2. It is recommended that 70% Isopropyl Alcohol should be applied to a soft clean cloth or tissue, not directly
on the component to be cleaned. The cloth should never be wet, just damp. A mild soapy water solution
is an alternative cleaning liquid.
3. To ensure that cross contamination does not occur, use clean cloth or sealed Alcohol swabs for each
device to be cleaned.
4. Wipe the surfaces of the Operator headset and headphone pads with the 70% Isopropyl Alcohol. Clean
other transducers in the same way.
Do not let 70% Isopropyl Alcohol or water enter the microphone sound
inlet.
5. Wipe the surfaces of the probe microphone, black ear loop and headphone pads with the 70% Isopropyl
Alcohol. Clean other transducers in the same way.
Do not let 70% Isopropyl Alcohol or water enter the
microphone sound inlet.
6. The probe microphone cords and white device housing may also be wiped with 70% Isopropyl Alcohol.
The speaker controls, headphone ear pads, head band and other components may be cleaned in a similar
way.
7. The white device housing may also be wiped with 70% Isopropyl Alcohol. The speaker controls,
headphone ear pads, head band and other components may be cleaned in a similar way.
8. Let all components that have been cleaned, thoroughly dry before use.
9. Cleaning of the computer should be performed using the methods suggested in the computer’s manual.