ARC-I-MARCI-2 MedRx AVANT ARC Installation Manual  

Effective Date: 10-21-2016

 

Getting to Know Your AVANT ARC 

 

The AVANT ARC (Audiometer and Real Ear Measurement Combination) represents a new era of ultra 
compact diagnostic audiometry and precision in-situ verification for your office. Compact yet rugged, this 
PC-Based system is USB powered and supports current ANSI and IEC audiometric and Real Ear tests. 

This Installation Manual and the Training Manual are supplied with your device in pdf format on a 
CD. Software to view the pdf files is included on the CD for your convenience. 

 

Intended Use Statement 

The MedRx AVANT ARC is an electronic instrument containing an audiometer intended to diagnose 
hearing loss in adults and children. Audiograms are created and used to set the correct gain levels of the 
hearing aid for various frequencies. The device also contains a REM (Real Ear Measurement) function 
intended to measure sound levels directly in patient ears. They are used in the measurement and fitting of 
hearing aids for adults and children. These measurements can be done with or without the hearing 
instrument being worn. They are also used for hearing loss simulation. The REM can perform Live Speech 
Mapping (MedRx’s in-situ method for getting the fitting right the first time). These devices should be 
operated by trained professionals with education and/or training in the field of audiometry. 

 

Indication For Use Statement 

The MedRx Avant ARC is an audiometer and a REM (Real Ear Measurement) device combined into one 
system. It is for use by professionals with education and/or training in the field of audiometry to conduct 
diagnostic hearing evaluations, evaluate basic hearing function, aid in the diagnosis of otologic disorders, 
as well as to evaluate the fitting of hearing aids and for hearing loss simulation in adults and children. 
 
The software includes targets for DSL v5.0 and NAL-NL2. A unique feature of the Real Ear Measurement 
System

 

is simultaneous, binaural Live Speech Mapping, MedRx’s in-situ method for getting the fitting right 

the 

first

 time, 

every

 time. The device performs Real Ear Measurements where measurements are 

performed directly on the client's ear. These measurements can be done with or without a hearing 
instrument being in place. The reason to perform Real Ear Measurements for hearing instrument fittings is 
that shape and size of an ear canal significantly influence a hearing instrument's performance. Knowing the 
sound intensity outside and inside the ear enables the practitioner to determine the actual amplification and 
output that the patient receives through the hearing aid and make the appropriate adjustments. 
 
This unit is a Type 2 AE audiometer that meets the requirements of both ANSI S3.6 and IEC 60645-1 and 
IEC 60645-2. These standards specify the required precision that the AVANT ARC does achieve. 
Measurement uncertainty is +/- 1 dB. This device and transducers are intended for use at test frequencies 
between 125 Hz and 8000 Hz.  This unit meets the requirements of both ANSI S3.46-1997 and IEC 
61669:2001. These standards specify the required precision that the Avant ARC does achieve.

 

 
The AVANT ARC is an electronic instrument intended for the testing of human hearing using both tone and 
speech audiometry. This device meets the specifications and tolerances for audiometers and standard 
reference threshold levels for audiometric transducers such as supra-aural, circumaural, and insert 
earphones, bone vibrators, and loudspeakers as defined in ANSI S-3.6 and IEC 60645. 
 
The AVANT ARC conforms to the RoHS directive 2011/65/EU when used with the supplied accessories. 

 

 

 

 

Summary of Contents for AVANT ARC

Page 1: ...INSTALLATION MANUAL www medrx usa com...

Page 2: ...alibration Files 14 EMC Precautions 15 Safety 18 Cleaning 19 Technical Information 20 Limited Warranty 23 0086 www medrx usa com 1200 Starkey Rd 105 Largo FL 33771 U S A Toll Free 888 392 1234 727 584...

Page 3: ...agnosis of otologic disorders as well as to evaluate the fitting of hearing aids and for hearing loss simulation in adults and children The software includes targets for DSL v5 0 and NAL NL2 A unique...

Page 4: ...Selected Bottom View A unique feature of the AVANT ARC is the light panel which indicates which ear is selected in the software When the unit is powered up and no ear is selected the light shines gree...

Page 5: ...re into the opening then remove the screw driver Be sure the wire is secure Repeat until all the speaker wires are secured then plug both connectors into the device Use passive speakers with 4 ohm imp...

Page 6: ...p program starts If the Setup program does not automatically start Press the Win R keys on the keyboard Type D Setup exe in the window where D is the drive letter assigned to the CD ROM drive on your...

Page 7: ...Select Accept Click Next 5 This screen sets the language and location choice Make a selection Click Next 6 This screen indicates the location of the program files The default location is recommended...

Page 8: ...Effective Date 10 21 2016 7 Installation of Program is ready to start To continue click Install To make changes click Back 8 Wait while the InstallShield Wizard Installs the AVANT Audiometer Click Ne...

Page 9: ...On the Setup screen choose Install Avant REM 2 Wait for the program to setup the InstallShield Wizard 3 This is the Welcome screen To continue click Next 4 Read the Software License Agreement This im...

Page 10: ...or most users If necessary this location can be changed To continue with the default settings click Next To change the location of the files advanced users or system administrators only click Change 6...

Page 11: ...lls the Avant REM program Click Next 9 When the installation is complete Click Finish Your AVANT ARC is USB powered Connect the USB cable from the AVANT ARC to your computer as shown below Wait for th...

Page 12: ...age 4 Click Audio Properties 5 In the Windows Sound control panel make sure the MedRx Audio Device is not set as default If it is default change this by clicking on your system audio device non MedRx...

Page 13: ...gured properly during driver installation the Audio Tab will appear like the image on the left If not use the pull down lists to adjust the settings to match the image 5 Open Audio B tab 6 When the au...

Page 14: ...a CD bearing the same serial number as your AVANT ARC device The final step before using your AVANT ARC to evaluate hearing is to load these device specific calibration files onto the computer used t...

Page 15: ...equipment can affect the Avant ARC Guidance and manufacturer s declaration electromagnetic emissions The Avant ARC is intended for use in electromagnetic environment specific below The customer or the...

Page 16: ...80 to 800 MHz d 2 33 P 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distan...

Page 17: ...wer of transmitter W Separation distance according to frequency of transmitter meters 150 kHz to 80 MHz d 1 17 P 80 MHz to 800 MHz d 1 17 P 800 MHz to 2 5 GHz d 2 33 P 0 01 0 12 0 12 0 233 0 1 0 37 0...

Page 18: ...om 0 C to 50 C All components with patient contact are made of bio compatible materials This device does not produce any adverse physiological effects Install the device as directed by this manual to...

Page 19: ...oth should never be wet just damp A mild soapy water solution is an alternative cleaning liquid 3 To ensure that cross contamination does not occur use clean cloth or sealed Alcohol swabs for each dev...

Page 20: ...xternal input is through the computer CD memory card Wave file Operator Microphone Modulation Pulse tone 0 25 0 5 s on time Warble tone 5 sinus frequency modulation repetition rate 5 Hz Patient Respon...

Page 21: ...re 20 Pa 125 45 0 250 25 5 500 11 5 750 8 0 1000 7 0 1500 6 5 2000 9 0 3000 10 0 4000 9 5 6000 15 5 8000 13 0 9000 13 0 10000 13 0 11200 13 0 12500 13 0 Speech 19 5 Calibration values Calibration val...

Page 22: ...ent should perform a subjective instrument check once a week The purpose of routine checking is to ensure as far as possible that the equipment is working properly that its calibration has not noticea...

Page 23: ...reater than 100 F and water liquid damage Repair or replacement of the system as provided under this warranty is the sole and exclusive remedy of the purchaser MedRx shall not be liable for any conseq...

Page 24: ...t Probe Microphone Left Probe Microphone Client Headphone Patient Response Button Patient Talk Back Microphone Bone Conductor Air Conduction Right Air Conduction Left USB Connection Left and Right Spe...