D-0126021-A
20
EMC Precautions
The Avant Hearing Instrument Test (HIT) Box needs special precautions regarding EMC and needs to be installed according to the following EMC information.
List of all cables and maximum lengths of cables, transducers and accessories:
Transducer / Accessories
Maximum Cable length
USB Cable
2.9 meters
Battery Simulators (All Versions)
0.3 meters
Microphones (All Version)
0.3 meters
Warnings!
•
The use of accessories, transducers and cables other than those specified, with the exception of transducers and cables sold by the manufacturer of the
Avant Hearing Instrument Test box as replacement parts for internal components, may result in increased emissions or decreased immunity of the Avant
Hearing Instrument Test Box.
•
The Avant Hearing Instrument Test Box should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the Avant
Hearing Instrument Test Box should be observed to verify normal operation in the configuration in which it will be used.
•
The Avant Hearing Instrument Test Box may be interfered with by other equipment, even if that other equipment complies with CISPR emission
requirements.
•
The Avant Hearing Instrument Test Box does not have a life supporting function.
•
Portable and mobile RF communications equipment can affect the Avant Hearing Instrument Test Box.