D-0126021-A
9
Side Tool Bar
1.
On the right side of the screen,
Hearing Aid Ear Selection is
displayed.
2.
Hearing Aid Serial Number and
Type of Hearing Aid.
3.
Test Set-up window.
4.
Stimulus Selection (Select Speech
Noise, ICRA, to evaluate the
performance of noise reduction
circuits).
5.
Precalibrate Function. This function allows you to calibrate the
reference level without the hearing aid turned on and then run the
tests without referencing. (Precalibrate can be used if it is suspected
that a high gain hearing aid is leaking sound thus affecting the
reference microphone).
Settings Auto Test Preferences and Procedure Options
1.
In the Test Set-Up Window, add
or remove the tests you want to
include in your Auto Test Battery
by clicking on the red X or green
check mark. The tests with the
green check mark will run when
clicking on the Start Auto Button.
2.
Click Start to run a single test
that is selected in the Test Set-up
Window. The test will finish after
the measurement is complete.
3.
Click MSC (Manual Signal Control) to start a single test in continuous
mode. The test will not finish by itself and will have to be stopped by
clicking Stop icon.
4.
Click Start Auto to run the selected tests in sequence.