D-0117969-B 01/2019
14
1.
Complete a careful otoscopic examination as instructed by the
manufacturer of your otoscope.
a.
For MedRx Video Otoscopes, refer to the Help system
within the software by pressing the
F1
key or Clicking
the
Help
icon or text menu.
2. Basic Diagnostics.
a.
Using your audiometer and NOAH, measure and
record the following values:
b. Pure Tone Air and Bone conduction thresholds.
c.
Narrow Band Noise or Pure Tone UCL at 500, 1000,
2000 and 4000 Hz.
d.
Other tests if needed in your practice or jurisdiction.
3. Position patient as described earlier in this manual.
4. Calibrate Probe Tube as described earlier in this manual.
5. Measure and place the probe tube in the ear canal. (Use chart
to the right).
6. Complete a Live Speech Mapping measurement
UNAIDED.
a. Microphone or file input
b. Peak Measures
c. 1/3 Octave resolution
7. Carefully place the hearing instrument in the ear without
disturbing the probe tube.
8. Complete a Live Speech Mapping measurement
AIDED
.
9. Using the On Top Mode, make any adjustments needed to the
hearing instrument so that:
a. The LSM curve for Quiet Speech (50 dB) is at the
bottom of the white modified speech spectrum (canoe).
b. The LSM curve for Conversational Speech (65 dB)
should be within the white modified speech spectrum.
c. The LSM curve for loud sounds (85 dB) (very loud
speech, clapping hands, etc.) is approximately 10 dB
below the UCL curve.
Summary of Contents for AVANT REM Speech+
Page 1: ......
Page 2: ...D 0117969 B 01 2019 2...
Page 5: ...D 0117969 B 01 2019 5 View Live Speech REM Help F1 MHA HLS Tympanometer...
Page 6: ...D 0117969 B 01 2019 6 Advanced Options General Tab Audio A Audio B Video Otoscope...
Page 7: ...D 0117969 B 01 2019 7 Tympanometer...
Page 8: ...D 0117969 B 01 2019 8 Speaker Calibration Tools Calibrate Speakers Start recommended 2...
Page 10: ...D 0117969 B 01 2019 10 Audio Air...
Page 11: ...D 0117969 B 01 2019 11 Options File Start Stop...
Page 12: ...D 0117969 B 01 2019 12 CD Options Show Targets F1 Help...
Page 13: ...D 0117969 B 01 2019 13...
Page 15: ...D 0117969 B 01 2019 15 Start Discard Delete Back Help Help Topics Display...
Page 17: ...D 0117969 B 01 2019 17 Pink Speech Byrne Speech ANSI...
Page 18: ...D 0117969 B 01 2019 18...
Page 19: ...D 0117969 B 01 2019 19...
Page 20: ...D 0117969 B 01 2019 20 START REAR OK START REAR OK...
Page 21: ...D 0117969 B 01 2019 21 START Coupler Response Stored OK...
Page 22: ...D 0117969 B 01 2019 22 Start button Options 1 24 octave...
Page 24: ...D 0117969 B 01 2019 24 M H A separately for each ear Binaural Monaural...
Page 25: ...D 0117969 B 01 2019 25 File CD Record F1 Help Normal Simulate track return Reset...
Page 26: ...D 0117969 B 01 2019 26...