DRYING CHAMBER

 

To  use  the 

Drying  Chamber

,  place  the  hearing  aid  (including 

earmold if BTE) on the pad and cover it with the vacuum cup. You 
can  place  two  ITE’s  or  one  BTE  with  mold in  the  chamber  at  a 
time. Put  the  WAND/CHAMBER  switch  to  the 

CHAMBER 

setting. 

Press the 

START 

switch. Press firmly on the vacuum cup to ensure a 

good  seal. The approximate vacuum level in the chamber will be 
indicated on the display panel.

 

As  the  pump  runs,  it  displaces  moisture  vapor  from  inside  the 
hearing aid. Make sure the chamber cup is centered on the pad. 
Again, it may be necessary to apply light pressure on the top of 
the cup to attain a proper vacuum. This is very important for the 
DRYING CHAMBER to work properly. A CHECK VACUUM message 
will appear if a proper seal is not attained. 

The pump will automatically turn off at the end of the cycle. A 
message “TOGGLE TO VENT” will appear on the display. Move the 
WAND/CHAMBER switch to the WAND position to release the 
vacuum. Remove the hearing aid(s) from the chamber upon 
completion.

 

The cycle can be ended prior to the 4 minute timer by pressing the 
STOP switch.  Then, the chamber can be vented as described above, 
or the re-started, with the time remaining on the display. 

Moving the WAND/CHAMBER switch to the WAND position during 
chamber operation will vent the chamber and turn the pump off.  
Press START to resume wand operation.

 

 

CHANGE DRYING CHAMBER FILTER

 

When significant discoloration is visible, change the Drying Chamber 
Filter. Using tweezers grab and pull filter to remove. Place two or three 
dots  of  Elmer's  All Purpose  Glue  on  the  edge of  new filter.  Use 
tweezers to insert the glued filter. NOTE: Too much glue will clog 
the fabric.

 

 

VACUUM WAND  

 

To activate the 

Vacuum Wand

, press the WAND/CHAMBER switch 

to the WAND setting, press the START switch.  The  unit  will  now 
provide  a  continuous  suction  to  the  Vacuum  Wand.  Use  the 
wand  to  vacuum  ear  wax  and  debris  from  the  receiver tube, 
microphone inlet, vent and battery compartment of  the aid. If the  

 

Summary of Contents for UltraVac+

Page 1: ...O P E R AT I N G I N S T R U C T I O N S HearingInstrumentRestorationSystem www medrx usa com ...

Page 2: ...As with all of our products MedRx is pleased to provide the top notch service and support that you have come to expect Our associates constantly strive to see that every effort is put forth in satisfying your needs and requests I can assure you we will not waiver from this objective as we move into the future We are aware of what has made us one of our industry s leading equipment suppliers QUALIT...

Page 3: ......

Page 4: ...commended Procedure 7 Drying Chamber 8 Change Drying Chamber Filter 8 Vacuum Wand 8 Pressure Wand 9 Care and Maintenance 9 Troubleshooting Guide 12 Limited Warranty 13 Specifications 14 www medrx usa com 1200 Starkey Rd 105 Largo FL 33771 Toll Free 888 392 1234 727 584 9600 Fax 727 584 9602 Email medrx medrx usa com ...

Page 5: ...lectrical shock or serious injury do not remove the white cover of this equipment Service should be performed by factory trained and authorized personnel BASIC OPERATION The UltraVac was designed to give dispensers and technicians a functional tool for repairing restoring and maintaining hearing aids It is easy to use which makes it ideal for those who may be just getting started and powerful enou...

Page 6: ...atient any concerns they have with the operation of the aid Counsel patient on what you will be doing to improve or correct the problem If possible test the condition of the battery and if it is weak or dead replace it If the aid is dead or damaged have the patient acknowledge this before proceeding Remove Visible Wax Using the cleaning tool kit provided with the UltraVac clean any large visible a...

Page 7: ...pear on the display Move the WAND CHAMBER switch to the WAND position to release the vacuum Remove the hearing aid s from the chamber upon completion The cycle can be ended prior to the 4 minute timer by pressing the STOP switch Then the chamber can be vented as described above or the re started with the time remaining on the display Moving the WAND CHAMBER switch to the WAND position during chamb...

Page 8: ...for cleaning vacuum tips is included in the enclosed tool kit The wand assembly can be removed from the coupling on the base unit by pushing in the retaining ring and gently pulling on the tubing To reinstall simply insert the tubing back into the coupler CARE AND MAINTENANCE The UltraVac has been engineered to provide years of trouble free service There are no field serviceable components inside ...

Page 9: ...first step before re use 1 Gather all equipment cleaning solutions and materials required to clean the patient care devices 2 WASH hands and put on gloves prior to cleaning the devices Personal protective equipment should be changed if torn or soiled Use protective eyewear during the cleaning procedure 3 Visible or gross soil present and or blood or body fluid spills must be removed prior to clean...

Page 10: ... the exterior with Alcohol Swabs Immerse the hose in a bath of alcohol for 10 to 20 minutes lowering one end first to remove all the air from the hose The drying cup and chamber bottom may be cleaned with Alcohol Swabs After air drying run the system for 10 minutes and one drying chamber cycle to remove any residual alcohol The drying chamber filter is located in the center of the black rubber cha...

Page 11: ...in panel Check to make sure filter is properly attached and not blocked Check to make sure tips on either hose are not blocked use the pressure function to clear tip Call for service Ensure that the switch is in the CHAMBER position Check to make sure vacuum cup does not leak around pad or has any cracks Call for service 888 392 1234 Caution To avoid the risk of electrical shock or serious injury ...

Page 12: ...haser MedRx shall not be liable for any consequential or incidental damages or for breach of any express or implied warranty Except to the extent of applicable law any implied warranty merchantability or fitness of this product is limited to the duration of this warranty MedRx will at its discretion service and repair out of warranty products at the purchaser s request charging for parts and labor...

Page 13: ... Volt 1 25 amp SB 5x20mm 220 Volt 2 50 amp SB 5x20mm Operating Temperature 0 to 50 C 32 to 120 F Weight Approx 8 5 lbs 5 kg Dimensions approx 10 5 w x 6 25 h x 12 d 26 6cm w x 16cm h x 30 4cm d Vacuum Tips 14 15 2 18 and 2 20 gauge w SafetyLok attachment Drying Chamber 4 minutes with auto shut off Cycle Time Vacuum Pressure Wand Continuous use ...

Page 14: ...TheUltraVac SystemIncludes UtraVac BaseUnit PressureWand withHose VacuumWand withHose Filter 6VacuumWandTips Drying ChamberFilter InstructionManual CleaningToolKit DryingChamber ...

Reviews: