Table 4-3
lists the specifications for the Battery Support System 2.
Table 4-2
Replacement Parts and Accessories
Description
Part Number
Battery Support System 2 Operating Instructions
3010515
Battery Support System 2 Service Manual
3011409
LIFEPAK NiCd Battery
3009376
LIFEPAK SLA Battery
3009378
FASTPAK 2 Battery
3009375
FASTPAK Battery (white)
9-10424-18
FASTPAK Battery (gold)
9-10424-19
AC Input Power Cord
803650
Battery Support System 2 Wall Mount Bracket
3010932
Table 4-3
Battery Support System 2 Specifications
*
* All specifications are at 20°C unless otherwise stated.
**
Catalog Number 99407-000018
Function
Description
General Specifications:
Dimensions
31.5 cm x 37.6 cm x 9.7 cm (12.4 inches x 14.8 inches x 3.8 inches)
Weight
8.2 kg (< 18 pounds) (excluding batteries)
Number of battery wells
3. Charging, conditioning, and shelf-life test capability in all battery wells.
Power requirements:
100/120/220/240 + 10%, 50/60 Hz + 3 Hz
Fuses
Two fuses in the power input module (5 x 20 mm T 250V, Low or High break
capacity) per IEC 127-2, sheet 1 or 2, such as Bussman GDC, Littelfuse
218, or Schurter FST.
100/120 Volt Input: 1.6 Amp (T, or Time-Lag)
220/240 Volt Input: 0.8 Amp (T, or Time-Lag)
Environmental Specifications:
Liquid Ingress
IPX0 per IEC529. Not protected against ingression of fluids. Indoor use only.
Altitude, operating
To 4,572m (15,000 feet)
Altitude, non-operating
To 5,500m (18,045 feet)
Humidity
0 to 95%
Temperature, operating
Temperature, storage
5° to 45°C (41° to 113°F) (commercial)
0° to 50°C (32° to 122°F) (military configuration)**
-40° to 70°C (-40° to 158°F) (commercial)
-46° to 71°C (-51° to 160°F) (military configuration)**
Vibration
IEC 68-2-6 (commercial)
Military Configuration:
**
USAF: MIL-STD-810F Method 514.5, Jet Aircraft (Figure 514.5C-6 and
Table 514.5C-VIII (General Exposure), 3.54 Grms, 30 min/axis) and
Turboprop (C130 curve, 4.87 Grms, 60 min/axis)
USAARL: MIL-STD-810E Method 514.4 (1.58 Grms, 60 min/axis)