115
Appendix
Electromagnetic Compatibility Compliance Information
In accordance with IEC 60601-1-2, the following information on electromagnetic compatibility
compliance has been provided. Also refer to the section “Interference from Electrical Equipment” in
Chapter 6: System Maintenance.
Table A1
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The Guardian RT is intended for use in the electromagnetic environment specified below. The
user of the Guardian RT should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The Guardian RT Monitor uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The Guardian RT is suitable for use in all
establishments, including domestic establishments,
and those directly connected to the public, low-
voltage, power-supply network that supplies
buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Not Applicable
Not Applicable
Voltage
Fluctuations/Flicker
Emissions
IEC 61000-3-3
Not Applicable
Not Applicable