Routine Operation
55
Entering Meter BG for Calibration
In order for the electronic signals from the Sensor to be converted into glucose values, Meter BG (Blood
Glucose) readings must be entered into the Guardian RT for the purpose of calibration. See Chapter 2:
Getting Started, Sensor Calibration.
From the HOME screen:
1.
Press the SEL button 2 times and the METER BG screen will appear. In the example below, the last
entered meter blood glucose value was 95 mg/dl (5.3 mmol/l). This value will thus appear on the
display. (If no BG reading was entered before, dashes (“- - -”) will appear on the screen.)
2.
Press the ACT button, and the number display will go to dashes. Enter the desired meter blood
glucose value, in increments of 1 mg/dl (0.1 mmol/l), by using the UP or Down arrow button.
Remember, the number value will start at 100 mg/dl. See Chapter 2: Getting Started, Sensor
Calibration.
NOTE -
If you scroll below 40 mg/dl (2.22 mmol/l) or above 400 mg/dl (22.2 mmol/l) the screen will
display “NO CAL.” The value entered will be stored but not used for calibration.
3.
Press the ACT button to set the value, and the display will return to the HOME screen.
MEter BG
95
mg/dl
12:00
AM
ENter BG
mg/dl
---
SET