70
Chapter 4
Low
NOTE -
If the alert pattern is set on beep, a series of descending tones will be heard. This tone is
unique to the LOW alert. If the alert type is set on vibrate, there is no differentiation between
the vibrations for the different alarms.
NOTE -
In the screen above, 60 mg/dl is being used as an example.
Error Message: LOW (Low Glucose Alarm)
Cause: Glucose value is lower than or equal to the low glucose limit set.
Action:
1.
Press the SEL button and then the ACT button to clear the alarm. The alarm will turn off and the
HOME screen will reappear. The alarm will repeat about every 20 minutes (or longer if the Repeat
interval has been customized) until the condition has been corrected.
2.
Take a blood glucose meter reading. If the value indicates low blood glucose level, treat per your
doctor’s instructions and enter the value into the Monitor for calibration.
3.
Even if the blood glucose reading does not confirm the alarm, enter the value into the Monitor for
calibration.
LOW
12:00
AM
mg/dl
GLUCOSE
60