Cal Err
Error Message: CAL ERR (Calibration Error)
Cause: Calibration cannot be done successfully due to Monitor and Transmitter not communicating,
Meter BG outside expected range from Sensor reading, or Monitor was turned off for more than
2 hours.
Action:
1.
Press the SEL button and then the ACT button to clear the alarm. The alarm will turn off, and the
HOME screen will appear. An alarm to recalibrate will repeat until a new calibration is
done.
2.
Check to make sure that the antenna icon is displayed. If it is not displayed, the communication link
between the Transmitter and the Monitor has not been set. Restore communication between the
Transmitter and the Monitor. Make sure the Monitor and Transmitter are within 6 feet (2 meters) of
each other.
3.
Take fingerstick reading and enter meter BG value, between 40-400 mg/dl (2.2-22.2 mmol/l), for
calibration.
CAL ERR
12:00
AM