85
Glucose Sensor Specifications
Model
: MD-‐JY-‐006
Storage Temperature Range
: +2°C ~+30°C
Storage Relative Humidity Range
: 20%~90%RH
Storage Atmospheric Pressure
: 700~1060 hPa
Glucose Range
: 2.2~22.2 mmol/L (40~400 mg/dL)
Sterilization Method
: By radiation
Sensor Life
: Up to 7 days
CGM System Accuracy
A multi-‐center, randomized clinic study is
designed to determine the sensor accuracy in
persons with Type 1 or Type 2 diabetes. In-‐clinic
testing consisted of frequent venous blood
sample testing (by Yellow Springs Instrument
2300 STAT Plus™ Glucose Analyzer, YSI) on a
random day in the 7-‐day sensor life. The
accuracy is based on the percentage of CGM
glucose readings that are within (±) 20%, 30% or
40% of YSI values at glucose values at or above