86
(>=) 75 mg/dL (4.2 mmol/L), or within 20 mg/dL
(1.1 mmol/l), 30 mg/dL (1.7 mmol/L) or 40 mg/dL
(2.2 mmol/L) of YSI values at glucose values
below (<) 75mg/dL (4.2 mmol/L).
Table. Percentage of CGM Glucose Readings
within ±20%/20, ±30%/30, or ±40%/40 of the YSI;
Calibrating every 12 hours, Abdomen insertion
site.
Number of
Matched
Pairs
CGM-‐YSI
±20%/20
±30%/30
±40%/40
1734
90%
96%
99%