Page 14 (18)
5171
31.07.13 TB
22.10.14 TB
LTR-1200
User Manual
3.3 Back Panel
Picture of 4 Terminal Version. (Only approx, the Sensor connectors are wrong)
Picture of the 8 Terminal Version:
By standard, the GND potentials of all built-in TEC and LDD devices are interconnected. Consequently,
all external loads and sensors to be attached to the back panel should be potential-free. At the most, only
one single potential (e.g. GND of one laser diode) may be tied to a given potential; it is however
recommended to keep all loads and sensors potential-free.
Upon customer request, individual outputs can be equipped such that they are potential-free (i.e.
independent of other built-in devices). Please contact Meerstetter Engineering with your enquiry
3.3.1 TEC Sensors Pinout
Fully pre-confectioned cables are available from Meerstetter Engineering at request.
3.3.1.1
Part List
Description
Manufacturer
Manufacturer Part Nr.
Connector Solder 7-Way
Binder
99-4925-00-07
Cable 6-Way, Shielded
Kabeltronik
2-LifYCY TP (820600800)
3.3.1.2
Pinout
Function
Pin Cable Color
Object Voltage Sense + 1
White
Object Voltage Sense -
2
Brown
Object Current Output
3
Green
Object Current Return
4
Yellow
Sink Sense A
5
Grey
Sink Sense B
6
Pink
NC
7
-
Please connect “Current Output” and “Voltage Sense +” together to one side of the thermistor,
and “Current Return” and “Voltage Sense –“ together to the other side of the thermistor.
Summary of Contents for LTR-1200