2  Functional Description 

2.1  Device-specific Documentation 

The following devices can be installed into the LTR-1200. Some of these devices can be connected in 
parallel to achieve higher output currents. Please contact Meerstetter Engineering if you have 
requirements which currently are not covered by any of these devices. 

LDD-Family 

(

L

aser 

D

iode 

D

rivers) 

TEC-Family

 (

T

hermo 

E

lectric 

C

ooling Controllers) 

  LDD-1121 

  LDD-1124 

  LDD-1125 

  TEC-1089 

  TEC-1090 

  TEC-1122 

  TEC-1123 

The functional description and the technical data of these devices are documented in their own specific 
Datasheets and User Manuals. 

2.2  Getting Started 

1.  Connect the LTR-1200 to the mains power and switch it on. 

  The display shows the start screen 

2.  Wait until the LTR-1200 has finished startup. 

  The display now shows the main screen 

  No error is indicated (both LEDs are green) 

3.  Use the navigation switches to browse through the Menu. The Menu structure is documented in 

chapter 2.3 Menu Structure 

4.  Connect to an internal device. (Chose one of the two options below): 

(For more information about how to connect please refer to 2.4 Communication Interfaces) 

  Ethernet (preferred, if possible): 

Remote Control over big distances. 
Simultaneous connections (multiple applications) to several devices are possible. 
Simultaneous connections (multiple PCs) to one device are possible. 

o

  Set up the IP configuration (See chapter 2.4.2 Connecting LTR-1200 over 

Ethernet) 

o

  Connect the Ethernet cable. 

  USB: 

Easy on-site connection. 
Only one device at the same time. 

o

  Connect the USB Cable. 

5.  Connect the corresponding (TEC or LDD) Service Software to a device  

(Notice that you establish a connection to a device. “Device” and “Terminal” have different 
meanings. See chapter 2.4.1 LTR-1200 Communication) 

  Set up the Service Software connection in the Maintenance Tab (See chapter 2.1 

Device-specific Documentation) 

  When the Software is successfully connected it will show a green status “Connected” and 

the Monitor tab will be updated periodically. 

  If you connect over Ethernet and wish to connect to multiple devices repeat this 

procedure for all devices to connect. 

6.  Now the device(s) are connected. The device configuration and the various setting options are 

described in the device-specific User Manuals. (See chapter 2.1 Device-specific Documentation) 

 
 
 

Summary of Contents for LTR-1200

Page 1: ...01 Email contact meerstetter ch Website www meerstetter ch Meerstetter Engineering GmbH ME reserves the right to make changes without further notice to the product described herein Information furnish...

Page 2: ......

Page 3: ...tware Default Address 9 2 4 4 Access over Third Party Software 9 2 4 5 Communication Fault Management 9 3 Technical Data 10 3 1 Elements and Configurations 10 3 2 Front Panel 10 3 2 1 LCD 10 3 2 2 Nav...

Page 4: ...ated interfaces therefore the originally supplied TEC and or LDD Service Software s can be used For continuous monitoring and control by customers systems the serial communication protocol MeCom can b...

Page 5: ...the two options below For more information about how to connect please refer to 2 4 Communication Interfaces Ethernet preferred if possible Remote Control over big distances Simultaneous connections m...

Page 6: ...us IP Gateway IP Network Mask FANC Settings Device Status T 1 Detailed T 1 Settings Object Temperature Device Overview Device Status Sink Temperature Actual Output Current T 2 Detailed T 2 Settings La...

Page 7: ...rvice Software Default Device Interface RS232 Baud Rate Interface RS485 Baud Rate HMI Device Address Internal Maximal Temperature Internal Maximal Fan Speed 2 3 1 2 LDD Laser Diode Current Actual Lase...

Page 8: ...represents only the output on the Back Plane of the LTR 1200 One device can be connected to more than one Terminal Channel represents the output channel of one device e g CH1 or CH2 in the case of TEC...

Page 9: ...fic LDD TEC Device Address in the Display menu In this case the Device Address 0 will be routed to the Service Software Default Device 2 4 4 Access over Third Party Software For Third Party Software c...

Page 10: ...etter Engineering with your enquiry Important note concerning the integration of LDD Family devices into LTR 1200 rack enclosures The internal wiring from an LDD device to its back panel terminal can...

Page 11: ...IN2 RES7 RES7 RES2 Input 3 IN3 RES6 RES6 RES3 Output 1 OUT RES5 RES5 RES4 GND Digital IO Ground All digital I Os are ESD protected and have a series and pull down resistor 3 2 4 2 Characteristic Symbo...

Page 12: ...t Characteristics Driver VOD1 Differential Output Voltage Loaded RL 100 RS422 2 0 3 6 V VOD2 Differential Output Voltage Loaded RL 54 RS485 1 5 3 6 V 3 2 6 RS232 3 2 6 1 Pinout Pin Function 2 TxD Data...

Page 13: ...ody Model 15 kV Electrical Characteristics Speed Communication Speed Baud Rate Software Setting 4800 1M bps 3 2 8 Ethernet 3 2 8 1 Specifications The Ethernet meets the following specifications IEEE 8...

Page 14: ...equipped such that they are potential free i e independent of other built in devices Please contact Meerstetter Engineering with your enquiry 3 3 1 TEC Sensors Pinout Fully pre confectioned cables ar...

Page 15: ...tetter Engineering at request 3 3 2 1 Part List Description Manufacturer Manufacturer Part Nr Connector Solder 4 Way Binder 99 4909 00 04 Cable 4 Way Shielded Kabeltronik 2 LifYCY TP 820400800 3 3 2 2...

Page 16: ...please refer to the LDD Family or TEC Family User Manuals Error Numbers 100 and above are HMI LTR specific 3 4 1 RTOS Errors Code Description Error Condition Remedy 80 RTOS_STACKOVERFLOW One task has...

Page 17: ...31 07 13 TB 22 10 14 SK Page 17 18 5171C 3 5 Physical Dimensions All dimensions are in mm Rack mounting brackets are removable 3 6 License Notice The LTR 1200 Software is built on FreeRTOS http www f...

Page 18: ...Comments Special Requirements Output 3 Output 4 TEC PT100 4 Wire _____A TEC PT100 4 Wire _____A PT1000 2 Wire _____V PT1000 2 Wire _____V NTC NTC LDD No LPC _____A LDD No LPC _____A LPC _____mA _____...

Reviews: