General Operation, Maintenance, and Service Information
Cleaning and Decontamination
Keep the unit clean and in a dry environment. To clean the unit, power down and unplug the instrument. Wipe with
a clean, water dampened cloth. Do not submerge in water or other cleaners or solvents. To clean the screen, take a
soft, water dampened cloth and gently wipe the surface.
Test Leads
Kelvin type high-voltage test leads rated for 16kV DC. The leads can be used on any ADX model.
Refer to the
Megger Baker ADX User Guide
for complete technical specifications on the unit and all accessories.
The test leads may be cleaned with a cloth dampened with water. Make sure the leads are dry before use.
Power Requirements
Using the provided AC power cord, connect the unit to a grounded AC power source.
The unit’s power requirements are 90–264 VAC, 47–63 Hz, 2.5 A AC maximum current draw.
A manually-reset circuit breaker protects the unit.
Environmental Conditions
The unit has been tested for use up to 3,000 meters (9,842 ft).
Only operate the tester in temperatures ranging from 5–40° C (41–104° F).
This unit is for use at a maximum relative humidity of 80% for temperatures up to 31° C (88° F), decreasing linearly
to 50% relative humidity at 40° C (104° F).
This unit is intended for Installation Category II in a Pollution Degree II environment.
www.megger.com/baker
Megger Baker ADX Quick Reference Guide
8
General Operating and Safety Information