Recommended Testing Sequence
To properly test Assets and to establish effective predictive maintenance programs, we suggest using a regimented
testing sequence. The approach involves conducting a series of progressively rigorous tests, and if a test fails,
troubleshooting and repairing at that time.
Testing should continue only after satisfactory diagnosis and repair.
The suggested testing sequence is:
1. Low-voltage RLC (resistance, inductance, capacitance)—balance resistance test.
2. Megohm/insulation Resistance (IR), dielectric absorption (DA), polarization index (PI).
3. HiPot (standard, ramp, or step).
4. Surge.
NOTE: This chapter includes brief descriptions of each test type conducted with the ADX analyzer.
For more details on the theory and application of each test, refer to the
Motor Testing Theory
Reference Manual
. For more details on each test type and the user interface, refer to the
Megger
Baker ADX User Guide.