DET2/3
21
Continuity Test
Manual mode
Continuous graphical mode
6
Press
Press
See Data management (page 33).
Note:
Press Save at any time to save the current reading.
Null test leads
Note:
Test must be running to be able to Null the test leads.
The Null function only works when the measured resistance is less than 10 Ω.
1. Put the two test lead tips firmly together.
2. Press
.
When the test result shows, press
again to activate / de-activate the Null process:
Null
active: Result is minus the test lead resistance.
The Null function is active while results are continuously updated or when results update is stopped.
Null
de-active: Result includes test lead resistance.
If the measured resistance is below zero, while Null is active, the result will show that it is too low to measure
(the instrument will not show negative resistance values).
OR