DET2/3
37
Maintenance
Battery replacement.
1
2
3
4
5
6
7
Warning: Disconnect all leads before removing case.
Caution: Protect the face and switches of the instrument when turning it upside-down.
Note:
Replace with Megger approved battery, part number 1002-552 only.
1. Remove lid by opening to approximately 70° and sliding right.
2. Switch the instrument
Off
.
3. Disconnect all leads .
4. Invert the lower case and rest the front panel on a soft surface so as not to damage the keypad.
5. Remove 4 fixing screws (1) and lift off the case.
6. Undo 4 fixing screws (2), remove battery cover (3).
7. Disconnect battery connector (4) and lift foam strips (7) to release the battery lead.
8. Remove battery (5).
9. Connect new battery to connector (5).
10. Install new battery ensuring correct orientation and that lead is correctly routed in slot (6). Secure lead using
foam strips (7).
Warning: Do not reconnect any test leads until the battery cover is fitted.
1. Refit battery cover (3) and secure using 4 screws (2) to torque 20 cNm.
2. Refit case and secure using 4 screws (1) to torque 40 cNm.
3. Turn instrument upright and refit lid.
Note:
The battery must be disposed of according to local environmental regulations.