DET2/3
40
Specifications
Instrument specifications
Display
Backlit, colour, WQVGA display 5.25 in
Operating temperature
From -10 to 40 °C (14 to 104 °F)
Operating humidity
90% RH max at 40 °C
Storage temperature
From -20 to 60 °C (4 to 140 °F)
Temperature coefficient
< ±0.1% per °C over operating Temp.
Environmental protection
IP54 operational (lid open)
IP65 storage (lid closed)
Altitude:
Up to 2000 m above sea-level.
Measurement connection rating CAT IV 300 V (clamp terminals not isolated from measurement terminals)
Measurement output rating
±50 V 1 A, ac or dc.
Power supply
Internal Li-ion battery (rechargeable / replaceable)
External 100 to 240 V ac, 50 - 60 Hz (with adaptor)
External 18V dc supply (65 W)
Battery life
10 h. max. (typically)
Battery charge time
Fast recharge to 50%, 3 h for 100%.
Ambient temperature
(battery charge)
0 - 40 ºC
Safety
Meets IEC 61010
EMC
Meets IEC 61326
Dimensions
315 x 285 x 181 mm
(13.8 x 11.2 x 7.1 in)
Weight
4.8 kg (10.6 lb)
PC data download
USB 2.0
Data Management
Internal 500 record storage
USB Host
Send data to pen drive
Resistivity calculation
Wenner
Schlumberger
Tests
2P, 3P, ART (Selective), 4P, Stakeless (clamp) modes
Real time clock
Yes
RoHS compliant
Yes
Instrument calibration check tool
Electrical specification
Resistance
25 Ω ±0.1%
Mechanical specification
Dimensions
60 x 55 x 25 mm (2.5 x 2.25 x 1 in)
Weight
0.1 kg (0.2 lb) approximately
www.calcert.com
sales@calcert.com
1.800.544.2843
0
5
10
15
20
25
30