Megger Limited
PWS / Nov 06
DET4TC/R Launch
17
The diagram below demonstrates the operation of
‘ART’
.
The ICLAMP has the ability to measure an electrode down to 5% of the total test
current generated by the instrument. In other words the electrode measured can
have a resistance up to 20 times that of the total system and still be measured.
Values higher than 20 times cannot be measured using
‘ART’
, and so you will need
to carry out a traditional 3-pole measurement. See over leaf.
Potential Probe (P)
Current
Probe (C)
Ground
Electrodes
Building earth
connection/s
I Total
I
Ie
I
e
Ie
Ie
Ie Test
>
I Total
20
X