meters and panel markings easily without
approaching the energized high-voltage
terminals.
CLEARANCES
1.
Be sure adequate clearances are maintained
between energized conductors and ground to
prevent arc over. Such accidental arc-overs
may create a safety hazard or damage the
equipment being tested. The position of the
exposed conductors with respect to ground
can often be maintained in place with clean,
dry nylon rope. The chart below indicates
some recommended clearances that will limit
the danger of static voltages being induced or
developed on nearby insulated objects,
including people.
TEST
VOLTAGE
DIRECT
AIR PATH
STRIKE
DISTANCE
CREEP
PATH
ALONG
NYLON
ROPE
MINIMUM
CLEARANCE
TO PREVENT
STATIC
CHARGING
30 kV
> 3.0 cm
(1.18 in)
> 8 cm
(3.15 in)
1.1 m
(43.3 in)
2.
After completion, make a careful review of
the test setup. Tests of the attached apparatus
can then be performed.
Summary of Contents for MIT30
Page 2: ......
Page 3: ...30kV DC Insulation Tester MIT30 Catalog No MIT30 Instruction Manual...
Page 6: ...AVTMMIT30 EN Rev E02 June 2009 ii M...
Page 12: ...M AVTMMIT30 EN Rev E02 June 2009 6 M...
Page 14: ...M AVTMMIT30 EN Rev E02 June 2009 8 M...
Page 20: ...M AVTMMIT30 EN Rev E02 June 2009 14 M...
Page 24: ...M AVTMMIT30 EN Rev E02 June 2009 18 M...
Page 34: ...M AVTMMIT30 EN Rev E02 June 2009 28 M...
Page 37: ...OPERATION TROUBLESHOOTING AVTMMIT30 EN Rev E02 June 2009 31 Troubleshooting Chart Figure 1...
Page 38: ...M AVTMMIT30 EN Rev E02 June 2009 32 M...
Page 40: ...M AVTMMIT30 EN Rev E02 June 2009 34 Figure 2...
Page 41: ...NOMENCLATURE AVTMMIT30 EN Rev E02 June 2009 35 This page intentionally left blank...
Page 43: ...NOMENCLATURE AVTMMIT30 EN Rev E02 June 2009 37 Figure 4...