AVTMMIT30-EN Rev E02 June 2009
24
INSPECTION AND
MAINTENANCE PROCEDURE
Proceed according to the steps below:
1.
Visually inspect the unit case, noting that
hinges and case locks function properly
2.
Check for breaks and or cracks in the case or
lid
3.
Check that the cable bag remains firmly
affixed to the lid
4.
Remove the cables from the bag and set them
aside.
5.
Take note of the air transport valve; that it
rotates smoothly and cinches down firmly.
6.
Note the condition of the carry handle and
case feet.
7.
Wipe the exterior of the case with a damp
cloth; then dry with a clean dry cloth.
Detergents and special cleaning agents are
neither required nor necessary.
8.
Retrieve the cable set aside earlier, inspect the
cables looking for damaged insulation or
frayed jackets. Pay special attention to the HV
output cable, check the condition of the
crocodile clip, and ensure that the spring
tension is sufficient and not damaged. The
Summary of Contents for MIT30
Page 2: ......
Page 3: ...30kV DC Insulation Tester MIT30 Catalog No MIT30 Instruction Manual...
Page 6: ...AVTMMIT30 EN Rev E02 June 2009 ii M...
Page 12: ...M AVTMMIT30 EN Rev E02 June 2009 6 M...
Page 14: ...M AVTMMIT30 EN Rev E02 June 2009 8 M...
Page 20: ...M AVTMMIT30 EN Rev E02 June 2009 14 M...
Page 24: ...M AVTMMIT30 EN Rev E02 June 2009 18 M...
Page 34: ...M AVTMMIT30 EN Rev E02 June 2009 28 M...
Page 37: ...OPERATION TROUBLESHOOTING AVTMMIT30 EN Rev E02 June 2009 31 Troubleshooting Chart Figure 1...
Page 38: ...M AVTMMIT30 EN Rev E02 June 2009 32 M...
Page 40: ...M AVTMMIT30 EN Rev E02 June 2009 34 Figure 2...
Page 41: ...NOMENCLATURE AVTMMIT30 EN Rev E02 June 2009 35 This page intentionally left blank...
Page 43: ...NOMENCLATURE AVTMMIT30 EN Rev E02 June 2009 37 Figure 4...