ROUTINE MAINTENCE
AVTMMIT30-EN Rev E02 June 2009
27
4.
Energize unit.
5.
Engage high voltage (DANGER! HIGH
VOLTAGE PRESENT!).
6.
Raise MIT30 output voltage until 300 µA is
displayed or the 30 kV limit is achieved.
7.
Read current measured through calibration
ammeter.
8.
Read current measured by on-board ammeter.
9.
If the difference between the calibration
ammeter and the on board ammeter is greater
than 1%, replace the control PCB.
10.
Disengage high voltage.
11.
De-energize unit.
RECOMMENDED CALIBRATION
INTERVAL IS 18 MONTHS.