background image

2

NOTE

THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS.

Users of this equipment and/or their employers are reminded that National Health and Safety Legislation requires them to carry out valid risk assessments of

all electrical work so as to identify potential sources of electrical danger and risk of electrical injury such as inadvertent short circuits. Where the assessments

show that the risk is significant then the use of fused test leads may be appropriate.

G

SAFETY WARNINGS

Safety Warning must be observed during use.

n

The circuit under test 

must

be switched off, de-energised, isolated and checked to be safe 

before

insulation test connections are made. 

Make sure

the

circuit is not re-energised whilst the instrument is connected.

n

Circuit connections 

must

not be touched during an insulation test.

n

After completing a test, capacitive circuits 

must

be completely discharged before disconnecting the test leads. Capacitive charges can be lethal.

n

Tested items should be firmly shorted out with a shorting link, after discharge, until required for use. This is to guard against any stored dielectric

absorption charge subsequently being released thereby raising the voltage to potentially dangerous levels.

n

The voltage indicator and automatic discharge features should be regarded as additional safety features and not a substitute for normal safe working

practice.

n

It is rare, but in certain circumstances, breakdown of the circuit under test may cause the instrument to terminate the test in an uncontrolled manner,

possibly causing a loss of display while the circuit remains energised. In this event, the unit 

must

be turned off and the circuit discharged manually. 

n

Test leads, including crocodile clips, must be in good order, clean and with no broken or cracked insulation.

n

The instrument should not be used if any part of it is damaged.

n

Insulation testing in wet weather conditions might be hazardous. It is recommended that this instrument is not used in these circumstances. If this is

unavoidable, the user must take all necessary precautions.

n

This instrument is not intrinsically safe and must not be used in hazardous atmospheres.

Summary of Contents for MIT510/2

Page 1: ...M MIT510 2 5kV Digital Insulation Tester USER MANUAL...

Page 2: ...d items should be firmly shorted out with a shorting link after discharge until required for use This is to guard against any stored dielectric absorption charge subsequently being released thereby ra...

Page 3: ...ndicate best practice Cleaning Disconnect the instrument and wipe it with a clean cloth slightly damped with soapy water or Isopropyl alcohol IPA CONTENTS 3 Safety warnings 2 Cleaning 3 Introduction 4...

Page 4: ...harge the battery after each use A completely exhausted battery will take 14 hours to recharge INTRODUCTION POWER LEAD AND CHARGING 4 The MIT510 2 is a compact microprocessor controlled high voltage d...

Page 5: ...t first will perform a self check during which Ini is displayed After Ini disappears the instrument is ready for use 3 Check voltage at terminals display to be sure the load is safe to energise 4 Set...

Page 6: ...played briefly When the test has stopped the display continues to show the voltage present on the test leads Pressing either test voltage s or t button will then display the test voltage immediately b...

Page 7: ...rminals marked and G These terminals are designed to accept only the test leads supplied Shutters across the terminals prevent accidental ingress of dirt and other objects Test lead plugs interlock wi...

Page 8: ...display shows the last measurement made until the timer or voltage test settings are changed or the test start stop button is pressed Analogue display This simulates an analogue meter movement to give...

Page 9: ...en the positive and negative terminals These leakages have a significant effect and can occur through the air itself If the guard lead is connected to ground then since the negative terminal is at the...

Page 10: ...rt circuit charge current 3 mA 5 kV Capacitor charge time 3 seconds per F at 3mA to 5 kV Capacitor discharge time 120 ms per F to discharge from 5000 V to 50 V Capacitance measurement 500 V minimum te...

Page 11: ...nA at all voltages Interference rejection 1 mA rms per 250 V test voltage up to a maximum of 2 mA Timer range Counts up to 99 minutes from start of test Test regimes Auto IR Lead set Three flexible si...

Page 12: ...insulated small clips 8101 183 3 x 10 m with medium insulated clips 1000 441 3 x 15 m with medium insulated clips 1000 442 3 x 3 m with large insulated clips 6220 811 3 x 10 m with large insulated cl...

Page 13: ...or your Megger products Returning your product to Megger UK and USA service centres 1 When an instrument requires recalibration or in the event of a repair being necessary a Returns Authorisation RA n...

Page 14: ...2 Canada T 1 416 298 9688 Canada only T 1 416 298 6770 F 1 416 298 0848 E casales megger com Megger 501 Crystal Paradise Mall Off Veera Desai Road Andheri w Mumbai 400053 Maharashtra India T 91 22 267...

Reviews: