If the power lead supplied is not suitable for your type of mains outlet, do

not use an adaptor. Always use a power lead fitted with the correct plug. 

NB

: A plug severed from the power lead is a hazard if plugged into a live

socket. Severed plugs must be destroyed immediately.

The instrument is fitted with a two-pin IEC60320 power inlet. Most power

leads are made with three-core cable so the ground connection will not

be used. 

Power lead cable colour coding:

International

USA

Earth/Ground

Yellow/Green

Green

Neutral

Blue

White

Live (Line)

Brown

Black

If using a fused plug ensure it is fitted with a 3 Amp fuse.

The instrument can be powered from 95 - 240 V rms, ±10% 50/60Hz.

With power connected the battery will charge as long as a test is not in

progress. The 

Power On/Off

button has a green LED which illuminates

when mains power is present. 

For best battery lifetime, charge the battery after each use. A completely

exhausted battery will take 14 hours to recharge.  

INTRODUCTION

POWER LEAD AND CHARGING

4

The MIT510/2 is a compact microprocessor controlled high voltage d.c.

insulation tester, powered by internal rechargeable battery or mains

supply. A comprehensive LCD display shows resistance, current,

capacitance, voltage and timer measurements digitally, with resistance

being shown in analogue as well.

Features include:

n

Resistance or current measurement- front panel selectable

n

Resistance measurement range - 10 k

to 15 T

n

Front panel selectable test voltages - 250 V, 500 V, 1000 V, 2500 V, 5000 V

n

Incremental timer (minutes & seconds) – measures during test

n

Load capacitance measurement – displayed at end of a test

n

LCD backlight – front panel selectable

n

Battery level/charging indicator

n

Mains (line) input presence LED

n

High voltage warning flashing LED shows when greater than 50 V

present on test leads

n

High voltage warning LCD "flash" shows when greater than 50 V 

present on test leads

n

8 hours continuous testing on full charge.

Summary of Contents for MIT510/2

Page 1: ...M MIT510 2 5kV Digital Insulation Tester USER MANUAL...

Page 2: ...d items should be firmly shorted out with a shorting link after discharge until required for use This is to guard against any stored dielectric absorption charge subsequently being released thereby ra...

Page 3: ...ndicate best practice Cleaning Disconnect the instrument and wipe it with a clean cloth slightly damped with soapy water or Isopropyl alcohol IPA CONTENTS 3 Safety warnings 2 Cleaning 3 Introduction 4...

Page 4: ...harge the battery after each use A completely exhausted battery will take 14 hours to recharge INTRODUCTION POWER LEAD AND CHARGING 4 The MIT510 2 is a compact microprocessor controlled high voltage d...

Page 5: ...t first will perform a self check during which Ini is displayed After Ini disappears the instrument is ready for use 3 Check voltage at terminals display to be sure the load is safe to energise 4 Set...

Page 6: ...played briefly When the test has stopped the display continues to show the voltage present on the test leads Pressing either test voltage s or t button will then display the test voltage immediately b...

Page 7: ...rminals marked and G These terminals are designed to accept only the test leads supplied Shutters across the terminals prevent accidental ingress of dirt and other objects Test lead plugs interlock wi...

Page 8: ...display shows the last measurement made until the timer or voltage test settings are changed or the test start stop button is pressed Analogue display This simulates an analogue meter movement to give...

Page 9: ...en the positive and negative terminals These leakages have a significant effect and can occur through the air itself If the guard lead is connected to ground then since the negative terminal is at the...

Page 10: ...rt circuit charge current 3 mA 5 kV Capacitor charge time 3 seconds per F at 3mA to 5 kV Capacitor discharge time 120 ms per F to discharge from 5000 V to 50 V Capacitance measurement 500 V minimum te...

Page 11: ...nA at all voltages Interference rejection 1 mA rms per 250 V test voltage up to a maximum of 2 mA Timer range Counts up to 99 minutes from start of test Test regimes Auto IR Lead set Three flexible si...

Page 12: ...insulated small clips 8101 183 3 x 10 m with medium insulated clips 1000 441 3 x 15 m with medium insulated clips 1000 442 3 x 3 m with large insulated clips 6220 811 3 x 10 m with large insulated cl...

Page 13: ...or your Megger products Returning your product to Megger UK and USA service centres 1 When an instrument requires recalibration or in the event of a repair being necessary a Returns Authorisation RA n...

Page 14: ...2 Canada T 1 416 298 9688 Canada only T 1 416 298 6770 F 1 416 298 0848 E casales megger com Megger 501 Crystal Paradise Mall Off Veera Desai Road Andheri w Mumbai 400053 Maharashtra India T 91 22 267...

Reviews: