background image

Power On/Off button

The instrument will only turn on if this button is pressed, held and then

released when the display responds. The instrument will not turn on if the

button is released before the display responds, or if the button is held

down for too long. This is a safety feature to prevent the instrument being

inadvertently turned on.

The instrument is turned off either by pressing the button again, or if the

instrument is running on the battery, by timing out after 10 minutes of

inactivity.

Upon switching the instrument on the display will first show ‘Ini’ while it

undergoes a self-checking routine. When Ini disappears, the instrument is

ready for use. 

Test voltage 

s

and

t

buttons

Using these buttons one of five test voltages can be selected: 250 V, 500 V, 

1 kV, 2.5 kV, and 5 kV. The selected voltage is shown on the display.

If there is an external voltage greater than 50 V on the test leads, the high

voltage warning indicators are flashed, and the display shows this voltage

instead. The instrument will not perform a test if this voltage is greater

than 50 V.

During a test the display shows the actual voltage on the test leads. If the

test voltage is changed during a test, the new test voltage will be displayed

briefly. 

When the test has stopped, the display continues to show the voltage

present on the test leads. Pressing either test voltage 

s

or 

t

button will

then display the test voltage immediately before the end of the test.

Test start / stop button

A test will only start if this button is pressed, held and then released as

soon as the red high voltage warning indicator LED lights. The LCD and

red LED high voltage warning indicators flash when the test starts.

A test will not start if the button is released before the red LED shows, or

pressed continuously for longer than 5 seconds. This is a safety feature to

prevent a test being started inadvertently.

The presence of a voltage greater than 50 V on the test leads is indicated

with flashing high voltage warning indicators. Testing is disabled if the

external voltage exceeds 50 V.

Testing will stop if the test start / stop button is pressed again.

When a test has finished, the instrument will discharge the load, which

may take some time. The operator must always check the load has been

discharged before touching the test leads.

/I button

Pressing this button toggles the digital display between the insulation

resistance and current. 

B

button

Pressing this button toggles the display backlight on and off.

High voltage warning LED

This is a red LED next to the TEST button on the front panel. The LED

flashes when the voltage on the test inputs exceeds 50 V.

6

Summary of Contents for MIT510/2

Page 1: ...M MIT510 2 5kV Digital Insulation Tester USER MANUAL...

Page 2: ...d items should be firmly shorted out with a shorting link after discharge until required for use This is to guard against any stored dielectric absorption charge subsequently being released thereby ra...

Page 3: ...ndicate best practice Cleaning Disconnect the instrument and wipe it with a clean cloth slightly damped with soapy water or Isopropyl alcohol IPA CONTENTS 3 Safety warnings 2 Cleaning 3 Introduction 4...

Page 4: ...harge the battery after each use A completely exhausted battery will take 14 hours to recharge INTRODUCTION POWER LEAD AND CHARGING 4 The MIT510 2 is a compact microprocessor controlled high voltage d...

Page 5: ...t first will perform a self check during which Ini is displayed After Ini disappears the instrument is ready for use 3 Check voltage at terminals display to be sure the load is safe to energise 4 Set...

Page 6: ...played briefly When the test has stopped the display continues to show the voltage present on the test leads Pressing either test voltage s or t button will then display the test voltage immediately b...

Page 7: ...rminals marked and G These terminals are designed to accept only the test leads supplied Shutters across the terminals prevent accidental ingress of dirt and other objects Test lead plugs interlock wi...

Page 8: ...display shows the last measurement made until the timer or voltage test settings are changed or the test start stop button is pressed Analogue display This simulates an analogue meter movement to give...

Page 9: ...en the positive and negative terminals These leakages have a significant effect and can occur through the air itself If the guard lead is connected to ground then since the negative terminal is at the...

Page 10: ...rt circuit charge current 3 mA 5 kV Capacitor charge time 3 seconds per F at 3mA to 5 kV Capacitor discharge time 120 ms per F to discharge from 5000 V to 50 V Capacitance measurement 500 V minimum te...

Page 11: ...nA at all voltages Interference rejection 1 mA rms per 250 V test voltage up to a maximum of 2 mA Timer range Counts up to 99 minutes from start of test Test regimes Auto IR Lead set Three flexible si...

Page 12: ...insulated small clips 8101 183 3 x 10 m with medium insulated clips 1000 441 3 x 15 m with medium insulated clips 1000 442 3 x 3 m with large insulated clips 6220 811 3 x 10 m with large insulated cl...

Page 13: ...or your Megger products Returning your product to Megger UK and USA service centres 1 When an instrument requires recalibration or in the event of a repair being necessary a Returns Authorisation RA n...

Page 14: ...2 Canada T 1 416 298 9688 Canada only T 1 416 298 6770 F 1 416 298 0848 E casales megger com Megger 501 Crystal Paradise Mall Off Veera Desai Road Andheri w Mumbai 400053 Maharashtra India T 91 22 267...

Reviews: