background image

across the insulation between the bare cable and the external sheathing

due to the presence of moisture or dirt. Where it is required to remove

the effect of this leakage, particularly at high testing voltages, a bare wire

may be bound tightly around the insulation and connected via the third

test lead to the guard terminal ‘G’.

The guard terminal is at the same potential as the negative terminal. Since

the leakage resistance is effectively in parallel with the resistance to be

measured, the use of the guard causes the current flowing through surface

leakage to be diverted from the measuring circuit. The instrument

therefore reads the leakage of the insulator, ignoring leakage across its

surface.

Factory programming socket:

This is a 9 pin socket under the hinged cover and is not for customer use.

B

button:

Pressing this button toggles the display backlight on and off.

"HV On" LED:

This is a red LED next to the 

TEST

button on the front panel. As long as

there is a voltage greater than 50 V, from an internal or external source, on

the test inputs, the LED flashes. 

"Charging Indicator" LED:

This is a green LED next to the 

Power On/Off

button on the front panel.

When mains power is present it shows green.

Test Terminals:

There are three test terminal , - and G. These terminals are

designed to accept only the test leads supplied. Shutters across the

terminals prevent accidental ingress of dirt and other objects. Test lead

plugs interlock with the shutters and are released by rotating the test lead

plug a quarter turn.

The Guard terminal is explained below and only used in cases where

surface leakage currents need to be eliminated. Most measurements use

just the + and – terminals. The instrument’s internal voltage generator

drives the + terminal with respect to the – terminal, current being

measured in the – terminal. 

The Guard Terminal:

For basic insulation tests and where there is little possibility of surface

leakage affecting the measurement, it is unnecessary to use the guard

terminal i.e. if the insulator is clean and there are unlikely to be any adverse

current paths. However in cable testing, there may be surface leakage paths

7

Summary of Contents for MIT510

Page 1: ...M MIT510 5kV Digital Insulation Tester USER MANUAL...

Page 2: ...ces breakdown of the circuit under test may cause a loss of display presence of HV on the test leads will still be shown by a flashing red LED If this happens stop the test immediately by pressing the...

Page 3: ...r lead and charging the batteries 4 Quick start 5 Instrument controls and indicators 6 Power On Off button 6 V and V buttons 6 Test button 6 I button 6 B button 7 HV On LED 7 Charging indicator LED 7...

Page 4: ...y lifetime charge the battery after each use A completely exhausted battery will take 14 hours to recharge INTRODUCTION POWER LEAD AND CHARGING 4 The MIT510 is a compact microprocessor controlled high...

Page 5: ...orm a self check during which Ini is displayed After Ini disappears the instrument is ready for use 3 Check voltage at terminals display to be sure the load is safe to energise 4 Set test voltage requ...

Page 6: ...nly If the test voltage is changed during a test the new test voltage will be displayed briefly before the display again shows the voltage on the test leads After a test has been stopped the display s...

Page 7: ...voltage greater than 50 V from an internal or external source on the test inputs the LED flashes Charging Indicator LED This is a green LED next to the Power On Off button on the front panel When main...

Page 8: ...f the last test remains on the display until another test is started Digital Display The digital display shows the resistance or current being measured during a test The I button toggles between the t...

Page 9: ...d Voltage input range 85 260 V 50 60Hz Battery life 6 hours continuous testing at 5 kV Test voltages 250 V 500 V 1000 V 2500 V 5000 V Accuracy 23 C 5 to 1 T 15 1 T to 7 5 T 20 5 T to 10 T 5 counts Gua...

Page 10: ...ominal test voltage at 100 M load Current measurement range 0 01 nA to 5 mA Current measurement accuracy 23 C 5 0 2 nA at all voltages Display Analogue Digital 3 digits Interference rejection 2mA rms...

Page 11: ...ncluded Accessories 3m lead set 8101 181 User guide on CD ROM 6172 929 Optional Accessories 3m lead set with straight jaw clamps 6220 797 15m lead set 8101 183 3m lead set insulated clamps Contact Meg...

Page 12: ...rument for Repair If it is necessary to retun an instrument for repair a returns Authorisation number must first be obtained by contacting one of the addresses shown You will be asked to provide key i...

Page 13: ......

Page 14: ...er com Megger Pty Limited Unit 26 9 Hudson Avenue Castle Hill Sydney NSW 2125 Australia T 61 0 2 9659 2005 F 61 0 2 9659 2201 E ausales megger com Megger Limited 110 Milner Avenue Unit 1 Scarborough O...

Reviews: