background image

8

Voltage at Terminals:

When not testing, the voltage displayed is the test voltage previously

selected. 

If there is an external voltage greater than 50 V present, this will be

displayed regardless of changes made to the test voltage. In this case the

instrument will not do a test, the 

TEST

button red LED and display

warning symbols will flash to warn of the hazard until the external voltage

becomes less than 50 V.

When testing, the voltage displayed is the voltage present at the test

terminals of the instrument.

If the test voltage is changed, by pressing either of the V

and V

buttons during a test, the new test voltage is displayed momentarily before

the voltage at the test terminals is displayed again.

Timer:

The timer shows minutes and seconds. At the start of a test the timer will

start from zero and at the end of a test it will stop. The duration of the last

test remains on the display until another test is started.

Digital Display:

The digital display shows the resistance or current being measured during

a test. The 

/ I

button toggles between the two. After a test the display

shows the last measurement made until V

, V

or 

TEST

buttons are

pressed.

Analogue display:

This simulates an analogue meter movement to give the user a better

"feel" for how a measurement is progressing. The analogue display shows

resistance only.

Battery Bar Graph:

This is a battery symbol on the LCD display comprising 4 pairs of

segments. The battery is monitored continuously when the instrument is

turned on the charge remaining in the battery is shown in the table below.

Fully charged battery

75% charged battery

50% charged battery

25% charged battery

Some testing still possible but battery may 

fail at any time

Symbol flashes when there is not enough 

charge to do a test and the instrument turns

itself off

When mains power is present the indicator shows the battery is being

charged by animating the segments of the bar graph.

Summary of Contents for MIT510

Page 1: ...M MIT510 5kV Digital Insulation Tester USER MANUAL...

Page 2: ...ces breakdown of the circuit under test may cause a loss of display presence of HV on the test leads will still be shown by a flashing red LED If this happens stop the test immediately by pressing the...

Page 3: ...r lead and charging the batteries 4 Quick start 5 Instrument controls and indicators 6 Power On Off button 6 V and V buttons 6 Test button 6 I button 6 B button 7 HV On LED 7 Charging indicator LED 7...

Page 4: ...y lifetime charge the battery after each use A completely exhausted battery will take 14 hours to recharge INTRODUCTION POWER LEAD AND CHARGING 4 The MIT510 is a compact microprocessor controlled high...

Page 5: ...orm a self check during which Ini is displayed After Ini disappears the instrument is ready for use 3 Check voltage at terminals display to be sure the load is safe to energise 4 Set test voltage requ...

Page 6: ...nly If the test voltage is changed during a test the new test voltage will be displayed briefly before the display again shows the voltage on the test leads After a test has been stopped the display s...

Page 7: ...voltage greater than 50 V from an internal or external source on the test inputs the LED flashes Charging Indicator LED This is a green LED next to the Power On Off button on the front panel When main...

Page 8: ...f the last test remains on the display until another test is started Digital Display The digital display shows the resistance or current being measured during a test The I button toggles between the t...

Page 9: ...d Voltage input range 85 260 V 50 60Hz Battery life 6 hours continuous testing at 5 kV Test voltages 250 V 500 V 1000 V 2500 V 5000 V Accuracy 23 C 5 to 1 T 15 1 T to 7 5 T 20 5 T to 10 T 5 counts Gua...

Page 10: ...ominal test voltage at 100 M load Current measurement range 0 01 nA to 5 mA Current measurement accuracy 23 C 5 0 2 nA at all voltages Display Analogue Digital 3 digits Interference rejection 2mA rms...

Page 11: ...ncluded Accessories 3m lead set 8101 181 User guide on CD ROM 6172 929 Optional Accessories 3m lead set with straight jaw clamps 6220 797 15m lead set 8101 183 3m lead set insulated clamps Contact Meg...

Page 12: ...rument for Repair If it is necessary to retun an instrument for repair a returns Authorisation number must first be obtained by contacting one of the addresses shown You will be asked to provide key i...

Page 13: ......

Page 14: ...er com Megger Pty Limited Unit 26 9 Hudson Avenue Castle Hill Sydney NSW 2125 Australia T 61 0 2 9659 2005 F 61 0 2 9659 2201 E ausales megger com Megger Limited 110 Milner Avenue Unit 1 Scarborough O...

Reviews: