Specification
11
Output voltage
0 - 60 kV rms
Test Frequency
61,8 Hz
Transformer rating
500 VA
Output disconnection
Within 1 ms of detection of breakdown
Power supply
115 V (99 - 132 V), 230 V (198 - 264 V) (switch selectable) 50/60 Hz, 80 VA
Fuses
500 mA (T) HBC to IEC127/5 for 220/240 V supply
1 A (T) HRC to IEC127/5 for 110/120 V supply
Mains power cord fused plug (when applicable): 3 Amp to BS1362
Operation
Semi automatic with 0,5, 2 or 3 kV/s selectable rate of voltage rise
Pause function with automatic 1 minute timer for Withstand testing
Display
0,0 - 60,0 kV LED digital display
HV on LED
Ramp rate LED
Measurement Accuracy
2% ± 3 digits
Controls
Start, Pause, Stop, Ramp Rate select and Power On/Off switch
Safety
The instrument meets the requirements of IEC 61010-1
EMC
In accordance with IEC 61326-1
Operational inaccuracies
Refer to www.megger.com
Temperature Range
Operating:
0 to 40 ºC
Storage:
-20 to +70 ºC
Humidity Range
Operating:
80% RH at 40 ºC
Storage:
93% RH at 40 ºC, 95% RH at 25 ºC
Dimensions
380mm (H) x 401mm (W) (w/o handles at 40mm each) x 260mm (D).
Weight
17 kg
Cleaning
Wipe disconnected test set with a clean cloth dampened with soapy water or Isopropyl Alcohol (IPA)