3.

     SAFETY 

WARNINGS

 

Safety warnings and precautions must be read and understood before the instrument is used. 

 

They must be observed during use.

 

The oil  test set must be properly earthed.

 

The test chamber must be kept scrupulously clean.

 

If the test chamber cover is cracked or damaged in any way the test set must not be used, but sent for repair to an authorised agent.

 

The chamber door hinges are fitted with earth contact springs. These must not be damaged or corroded.

 

The correct type of oil test vessel must be used, and correctly positioned on the support horns before carrying out any testing.

 

Replacement fuses must be of the correct type and rating.

NOTE

THE TEST SET MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS.

TO PREVENT DAMAGE DO NOT TRANSPORT INSTRUMENT WITH TEST CELL FITTED IN TEST CHAMBER.

WEEE DIRECTIVE

The crossed out wheeled bin symbol placed on Megger products is a reminder not to dispose of the product at the end of its life with general waste.

Megger is registered in the UK as a Producer of Electrical and Electronic Equipment. The Registration No is WEE/HE0146QT

For further information about disposal of the product consut your local Megger company or distributer or visit your local Megger website.

Symbols used on the Test Set

 
     

G

 

Caution: Refer to accompanying notes.

   

     

F

 

Risk of electric shock.

     

c

 

Equipment complies with relevant EU Directives

Summary of Contents for OTS60SX/2

Page 1: ...M OTS60SX 2 Semi Automatic Oil Test Set User Guide...

Page 2: ...ampling 5 Preparing the Oil Test Vessel 5 Control Panel 6 Breakdown Testing 6 Withstand Testing 6 Warning messages 7 Testing to National Standards 7 Sampling and Testing Oil for Dielectric strength 8...

Page 3: ...sting Replacement fuses must be of the correct type and rating NOTE THE TEST SET MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS TO PREVENT DAMAGE DO NOT TRANSPORT INSTRUMENT WITH TEST CEL...

Page 4: ...sted according to breakdown specification by using a suitable sequence of tests The instrument is constructed in a strong sheet steel case Handles are provided for ease of transportation A pouch is lo...

Page 5: ...f the oil may lead to misleading results Always run off a small amount of oil before sampling and ensure that the flow of oil is steady Preparing the Oil Test Vessel Optional vessels are available for...

Page 6: ...left on the display until the next test is started If no breakdown occurs the high voltage will rise to 60 kV At this point the test voltage is cut off and this maximum value left on the display A tes...

Page 7: ...60SX 2 Oil Test Set The test vessel must be kept scrupulously clean Before each use dismantle the test vessel clean and dry re assemble and set the electrode gap to 2 54 mm 0 100 inch Dismantling the...

Page 8: ...l and the sampling apparatus as the condition of the oil itself After cleaning the apparatus it should under no circumstances be wiped dried or even handled with a dusty or fibrous cloth as loose dust...

Page 9: ...revent contamination is necessary when a test is to be made on a bulk delivery tanker and on all occasions when a high value of dielectric strength is anticipated The procedure suggested above should...

Page 10: ...3 min 3 min 5 min 5 min 3 min 3 min 3 min 3 min 3 min 3 min 3 min 3 min 10 min 3 min Rate of rise of voltage 2 kV s 3 kV s 0 5 kV s 0 5 kV s 2 kV s 2 kV s 2 kV s 2 kV s 2 kV s 2 kV s 2 kV s 3 kV s 3...

Page 11: ...n with automatic 1 minute timer for Withstand testing Display 0 0 60 0 kV LED digital display HV on LED Ramp rate LED Measurement Accuracy 2 3 digits Controls Start Pause Stop Ramp Rate select and Pow...

Page 12: ...Accessories Carrying Case 6420 106 Test Vessels 0 5 l for IEC156 spherical electrodes 6320 233 0 5 l for IEC156 with stirrer spherical electrodes 6320 236 0 5 l for VDE0370 mushroom electrodes 6320 2...

Page 13: ...contact Megger LImited or Megger Archcliffe Road Valley Forge Corporate Center Dover 2621 Van Buren Avenue Kent CT17 9EN Norristown PA 19403 England U S A Tel 44 0 1304 502243 Tel 1 610 676 8579 Fax...

Page 14: ...333 3201 T 33 1 30 16 08 90 F 1 214 331 7399 F 33 1 34 61 23 77 OTHER TECHNICAL SALES OFFICES Toronto CANADA Sydney AUSTRALIA Madrid SPAIN Mumbai INDIA and the Kingdom of BAHRAIN Megger products are...

Reviews: