G
SafETy WarninGS
Safety warnings and precautions
must
be read and under-
stood before the OTS80PB/OTS60PB is used. They
must
be observed during use.
n
The OTS80PB/OTS60PB
must
only be used in the way
intended and for the stated purposes described by
Megger.
n
The OTS80PB/OTS60PB
must not
be used if any part
of it is damaged.
n
The OTS80PB/OTS60PB
must
be properly earthed. A
supplementary earth terminal is provided.
n
Position the OTS80PB/OTS60PB so that power can be
easily removed.
n
The OTS80PB/OTS60PB
must not
be used in wet
locations or with rain falling on the instrument.
n
The OTS80PB/OTS60PB
must
only be used for testing
electrical grades of insulating liquids. No other objects
or substances should be placed in the test chamber.
n
The OTS80PB/OTS60PB
must
only be used with one
of Megger’s precision test vessels or VCM100D
correctly fitted on the support horns in the test
chamber.
n
Replacement fuses
must
be of the correct type and
rating.
n
If batteries are fitted, the 12 VDC charger
must
only
be powered from a negative earthed vehicle 12 VDC
source using the Megger supplied lead.
n
Calibration or repair
must
only be carried out by a
Megger qualified repair facility.
n
The test chamber
must
be kept clean and free from lint
or other fibres.
n
The OTS80PB/OTS60PB weighs between 16 kg and
23 kg. Care should be taken when lifting the
instrument.
n
It is prohibited to insert foreign objects into any gap on
the instrument before or during a test.
n
Unit must be used in a well ventilated location. Small
quantities of Ozone may be created after prolonged
testing.
Symbols used on this instrument
G
Caution: refer to accompanying notes
F
Danger high voltage, risk of electrical shock
DC voltage
a
AC voltage
c
Equipment complies with current EU
Directives
u
Earth terminal
f
Fuse
USB termina
l
Equipment complies with ‘C tick’
requirements
Do not dispose of in the waste stream
instrument setup
The OTS80PB/OTS60PB
must
only be used by suitably
trained persons.
Out of the box
List of items
n
OTS80PB/OTS60PB Oil test set instrument
n
400 ml precision vessel assembly
n
Electrode set(s) as ordered & feeler gauge set
n
Impeller stirrer lid assembly if ordered
n
Magnetic bead stirrers (2) and retriever
n
User manual CD
n
Power lead as ordered
n
Carry case with shoulder strap if ordered
accessories (supplied separately)
n
Vessel assembly 150 ml
n
Transportation case
n
VCM100D voltage meter