Page 28
Technical Specification
Electrical specifications
AC Voltage input range:
5 kV, 10 kV: 90-264 V rms 47 – 63 Hz 100 VA
15 kV:
90-264 V rms 47 – 63 Hz 200 VA
Battery:
11.1 V, 5.2 Ah meets IEC 62133:2003
Battery life
S1-586:
6 hours (typical) continuous testing at 5 kV with a
100 M
Ω
load
S1-1068:
4.5 hours (typical) continuous testing at 10 kV with
a 100 M
Ω
load
S1-1568:
4.5 hours (typical) continuous testing at 15 kV with
a 100 M
Ω
load
30 min. chg: 1 hour operation at 5 kV, 100 M
Ω
Test voltages: 250V, 500V, 1000 V, 2500 V, 5000 V, 10000 V,
15000, V
L
Lock test:
40 V to 1 kV in 10 V steps, 1 kV to 5 kV in 25 V
steps, 5 kV to 10 kV in 25 V steps
Test voltage
accuracy:
+4%, -0%, ±10 V nominal test voltage at 1G
Ω
load
(0 °C to 30 °C)
Resistance range:
S1-568 :
10 k
Ω
to 15 T
Ω
S1-1068:
10 k
Ω
to 35 T
Ω
S1-1568:
10 k
Ω
to 35 T
Ω
Accuracy (23 °C):
S1-568
5000 V 2500 V 1000 V 500 V 250 V
±5% to
1 T
Ω
500 G
Ω
200 G
Ω
100 G
Ω
50 G
Ω
±20% to
10 T
Ω
5 T
Ω
2 T
Ω
1 T
Ω
500 G
Ω
S1-1068
10000 V 5000 V 2500 V 1000 V 500 V
±5% to
2 T
Ω
1 T
Ω
500 G
Ω
200 G
Ω
100 G
Ω
±20% to
20 T
Ω
10 T
Ω
5 T
Ω
2 T
Ω
1 T
Ω
S1-1568
15000 V 10000 V 5000 V 2500 V 1000 V
±5% to
3 T
Ω
2 T
Ω
1 T
Ω
500 G
Ω
200 G
Ω
±20% to
30 T
Ω
20 T
Ω
10 T
Ω
5 T
Ω
2 T
Ω
Guard terminal
performance
:
Guards out parallel leakage resistance down
to 250k
Ω
with a maximum additional
resistance error of 1% with a 100M
Ω
load*
* Assumes that the guard ring is positioned centrally across the
insulating surface being guarded such that the resistances to
the test conductors are equal
Display range
analogue:
100 k
Ω
to 10 T
Ω
Display range
digital:
10 k
Ω
to 35 T
Ω
Short circuit curent : 6 mA nominal
Insulation alarm:
100 k
Ω
to 10 G
Ω
Capacitor chg bat pwr: < 2.5 s/µF to 5 kV, < 5 s/µF to 10 kV
< 6.3 s/µF to 15 kV
Capacitor chg AC pwr: < 1.5 s/µF to 5 kV , < 2.7 s/µF to 10 kV,
< 4.4 s/µF to 15 kV